9
Ergostyle ™ FX Tables
4.5 POWER ON/OFF TEST
A. Specification
Unit turns Off and On
B. Equipment Required
None
C. On/Off Switch Test Procedure
1.
Connect the Mains Power Cord into grounded
power source with appropriate voltage.
See Specifications Section. Unit should
come
on.
2.
Turn unit Off by disconnecting Mains Power
Cord from the grounded power source.
D. Test Result
1.
If unit turns Off and On, unit passed test.
2.
If unit does not turn Off and On, unit failed
test. Contact Chattanooga Group to obtain
Factory
Service.
4 TROUBLESHOOTING
4.6 PATIENT INTERRUPT SWITCH TEST
A. Specification
Table stops when Patient Interrupt Switch is
depressed.
B. Equipment Required
None
C. Patient Interrupt Switch Test Procedure
1.
Connect Mains Power Cord into grounded
power source with appropriate voltage.
2.
Connect Patient Interrupt Switch to unit
making certain it is completely seated in the
Patient Interrupt Switch receptacle. See
Figure
2.1,
item
26.
3.
Set up a treatment, refer to User Manual, and
press the START/STOP button.
4.
Press the Patient Interrupt Switch to stop the
treatment.
D. Test Result
1.
If table stops when the Patient Interrupt
Switch is depressed, unit passed test.
2.
If unit does not stop, repeat
test with a known good Patient Interrupt
Switch.
3.
If known good Patient Interrupt Switch
resolves the problem, replace existing Patient
Interrupt Switch with new part.
4.
If the known good Patient Interrupt Switch
does not remedy the problem, contact
Chattanooga Group to obtain Factory Service.
4.7 AUTO FLEXION AND AUTO DROP TESTS
If the table is equipped with either or both of
these options, verify that both features are
operational by following the instructions below.
A. Auto Flexion
Adjust angles and speed settings on Auto
Flexion to assure the Range of Motion and
speeds vary accordingly.
B. Auto Cocking
Verify that Auto Cocking funtions. If not,
check to ensure that the hoses are not pinched
or
kinked.
Содержание ERGOSTYLE FX 5820
Страница 24: ...Ergostyle FXTables 22 8 SCHEMATICS Block Diagram 1 of 1 ...
Страница 25: ...Ergostyle FXTables 23 8 SCHEMATICS Pneumatic 1 of 1 ...
Страница 26: ...Ergostyle FXTables 24 8 SCHEMATICS Driver Board 58432 1 of 1 ...
Страница 27: ...Ergostyle FXTables 25 8 SCHEMATICS Control Board 58433 1 of 1 ...
Страница 28: ...Ergostyle FXTables 26 8 SCHEMATICS Daughter Board 58543 1 of 1 ...