4- TROUBLESHOOTING
4.3
ELECTRICAL SAFETY
This table has been tested to UL 60601-1,
Standard for Safety for Medical Equipment.
A.
Power
Requirements
Models: 8080 and 8082 . . . . . . . . . . . . Input : 120 VAC
60 Hz, 5A
B.
Test Specifications
1. Equipment Required
as per UL 60601-1
NOTE:
Any indicating device (ohmmeter, battery-
and-buzzer combination or the like) may be
used to determine compliance with the
grounding continuity requirements in DVB.2.1
and DVE.2.1.1 of UL 60601-1.
As per
UL 60601-1. . . . . . . . . . . .Achieve Continuity
In accordance with UL 60601-1 Method A
Normal Conditiions . . . . . . . . . . . . . . . . . . . . <300 μA
Single Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . <600 μA
NOTE:
Ground Continuity and Leakage tests are
performed to verify the table meets the
requirements listed.
Test methods must comply with UL 60601-1.
Continuity testing should occur between the
controller mounting plate and the ground/earth
wire of the line cord or ground terminal.
Leakage testing should be performed on the
protective, earth or ground conductor, the
enclosures and any devices that are patient -
connected.
Listed by Intertek Testing Services NA
Inc.
Conforms to UL Standard 60601-1
Certified to CAN/CSA Standard C22.2
No. 601.1-M90 w/A2
Meets Directive 93/42/
EEC. Compliant to IEC/EN
60601-1, IEC 60601-1-2
9700675
A table that is failing the dielectric withstanding
and/ or leakage test could indicate serious
internal system problems. Do not place this
table back into service. Send the table to the
factory for repair. Do not attempt to repair the
table in the field.
4.4 GROUND CONTINUITY
4.6 LEAKAGE
4.5 DIELECTRIC WITHSTAND (HI- POT TEST)
NOTE:
NOTE:
This device complies with current leakage,
This device complies with current leakage,
ground continuity, and dielectric withstand
ground continuity, and dielectric withstand
(Hi-Pot) limits as prescribed by IEC/EN/UL
(Hi-Pot) limits as prescribed by IEC/EN/UL
60601-01 and CSA/CAN 301.1 Medical Electrical,
60601-01 and CSA/CAN 301.1 Medical Electrical,
Part 1: General Requirements for Safety.
Part 1: General Requirements for Safety.
Facility, local and national limits and test
Facility, local and national limits and test
methods may vary.
methods may vary.