5
Mōveo™ XP
Equipment not suitable for use in the
presence of a flammable anesthetic
mixture with air, oxygen, or nitrous
oxide.
Do not use with Oxygen tents. Use
oxygen administering equipment of the
nasal, mask, or ventilator type only.
This table should only be operated
under the prescription and supervision
of a licensed medical practitioner
that is familiar with the precautionary
measures and operational functions
associated with the table being used.
DO NOT connect the unit to an
electrical supply without first verifying
that the power supply is the correct
voltage. Incorrect voltage may cause
unit damage, malfunction, electrical
shock, fire, or personal injury. Your
unit was constructed to operate only
on the electrical voltage specified on
the Voltage Rating and Serial Number
Plate. Contact your Chattanooga Group
dealer if the unit is not properly rated.
NiMH Batteries contain Class E
Corrosive materials. In the event of
battery cell rupture or leakage, handle
battery housing wearing neoprene or
natural rubber gloves. Contents of a
ruptured or leaking battery can cause
respiratory irritation. Hypersensitivity
to nickel can cause allergic pulmonary
asthma. Contents of cell coming
in contact with skin can cause skin
irritation and/or chemical burns.
Use the Battery Backup only with the
Mōveo™ XP
.
•
•
•
•
•
•
Never, under any circumstances, open the
battery housing or cells. Should an individual
cell from a battery become disassembled,
spontaneous combustion of the negative
electrode is possible. There can be a delay
between exposure to air and spontaneous
combustion.
Charge the Battery Backup according to
the instructions found in this manual. Never
attempt to charge the Battery Backup on
any other charging mechanism.
Do not reverse the polarity of the Battery
Backup. Doing so can increase the individual
cell temperature and cause cell rupture or
leakage.
Dispose of NiMH batteries according
to national, state and local codes and
regulations.
Never dispose of Battery Backup in fire.
Never short circuit the battery. The battery
may explode, ignite, leak or get hot causing
serious personal injury.
Do not move or elevate any cushion section
while the patient’s weight is on that section.
Do not reposition or allow the patient to
get on or off table while the table is in any
position other than completely horizontal.
Do not allow anything to be under the table
while the table is descending.
Do not make any adjustment to the table
while the table is in any position other than
completely horizontal.
•
•
•
•
•
•
•
•
•
1.1 PRECAUTIONARY INSTRUCTIONS (CONTINUED)
1- SAFETY PRECAUTIONS