Chandler Engineering
5000 SERIES PUMP MANUAL
[74]
2
22 mm open
–end
wrench
The 22 mm open-end wrench fits the
flat area of the cylinder barrel. It is
used to hold the barrel in place while
tightening the fluid fittings, or while
tightening
the
cylinder
barrel
retaining ring.
3
Adjustable face
spanner wrench
The adjustable face spanner wrench
is used to loosen and tighten the
cylinder barrel retaining ring. It is also
used to remove and install the seal
retaining nut or wash ring assembly,
if your system has one.
4
Seal extraction rod
The seal extraction rod is a smooth
metal rod with the same outside
diameter as the piston. It is used to
extract a seal without damaging the
cylinder.
5
Seal guide ring
This ring guides the seal insertion
plunger into the cylinder during seal
replacement.
6
Seal insertion plunger
This is a 3
–step cylindrical rod used
to push a seal out of the seal guide
ring and into the cylinder barrel.
7
27 mm socket on
extension tube
This socket is inserted into the
cylinder housing to engage the piston
nut
during
piston
removal
or
replacement.
8
Speedbite
®
plugs and
nuts (2 each)
These parts are used to close off the
fluid ports during seal extraction.
Table 11-2
Spare Parts Kit
Item No.
Spare Part
1
Seals
Depending on the type of pump system that is ordered, the seals
included are one of the following:
•
Cup
–type ambient temperature seals made of ultra–high molecular
weight polyethylene (white UHMW P
–E)
•
T
–type high temperature seals made of Aflas
®
(black) with
accompanying Teflon
®
back
–up ring.
Содержание Ametek Quizix 5000 Series
Страница 8: ...Chandler Engineering 5000 SERIES PUMP MANUAL 6 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 14: ...Chandler Engineering 5000 SERIES PUMP MANUAL 12 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 18: ...Chandler Engineering 5000 SERIES PUMP MANUAL 16 Stand Mounted Q5200 System...
Страница 19: ...Chapter 2 SYSTEM SETUP 17 Expanded Front View of a Stand Mounted Q5200 System...
Страница 20: ...Chandler Engineering 5000 SERIES PUMP MANUAL 18 Stand Mounted Q5400 System...
Страница 34: ...Chandler Engineering 5000 SERIES PUMP MANUAL 32 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 40: ...Chandler Engineering 5000 SERIES PUMP MANUAL 38 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 48: ...Chandler Engineering 5000 SERIES PUMP MANUAL 46 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 66: ...Chandler Engineering 5000 SERIES PUMP MANUAL 64 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 70: ...Chandler Engineering 5000 SERIES PUMP MANUAL 68 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 120: ...Chandler Engineering 5000 SERIES PUMP MANUAL 118 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 126: ...Chandler Engineering 5000 SERIES PUMP MANUAL 124 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 127: ...125 Appendix A A Declaration of Conformity is found on the following page...
Страница 152: ...Chandler Engineering 5000 SERIES PUMP MANUAL 150 THIS PAGE INTENTIONALLY LEFT BLANK...