Damping Rate (Response Rate):
there are three optional items, i.e. slow speed,
medium speed and fast speed.
Slow Speed:
the response rate is slow and measuring accuracy is high, not easy to
be disturbed;
Medium Speed:
the parameters are between those for slow speed and fast speed;
Fast Speed:
the response rate is fast and measuring accuracy is low, very easy to be
disturbed. The factory default is medium speed.
Safe Level (Safe Material Level):
there are four optional items, i.e. remaining,
minimum value, maximum value and setting value.
Remaining:
the display value is the final measured value before wave loss, and the
current is the corresponding value;
Minimum:
the display value is 4mA after wave loss, and the current 4mA;
Maximum:
the display value is 20mA after wave loss, and the current 20mA;
Setting Value:
the display value is the final measured value after wave loss, and the
current output is the setting value of setting current. The factory default is the re-
maining value.
Current Set (Set Current):
set the output current after wave loss, which shall be
more than 3.6mA and less than 22mA, and become invalid in case of remaining/
minimum/maximum values on reselection. The factory default is 3.6mA.