(3) Arch Top Tank
For open container, the support shall be used for installation. The bearing capacity of
support shall be noted and certain distance shall be kept between the sensor and
container wall. If the upper part and lower part of the open container or stock bin in-
ner wall are flat and free of hanging objects and any other objects, the distance
between the sensor and container wall is detailed as follows:
Maximum
Measuring
Range
Minimum
Distance to
Wall
Maximum
Measuring
Range
Minimum
Distance to
Wall
Maximum
Measuring
Range
Minimum
Distance to
Wall
5m
0.6m
10m
1.2m
15m
1.8m
20m
2.5m
30m
3.5m
40m
5m
50m
6m
60m
7m
70m
8m
Installation on open container –
with top against the support at
one side
As the open container has no focusing effect, the sensor can be installed in the
middle of the container.
Installation on open container –
with support at the top middle
part