Chapter IV Spectrum Analyzer Mode
78
Generator
Generator
Off
On
Mode
CW
Track
Power
0dBm
CW Freq
1.000GHz
Trans Meas
Power Offset
0.0dB
Freq Offset
0.000Hz
< Back
·[Generator Off On]:
Press
【
Measure
】
→[More 1 of 2]]→[Generator]→[Generator
Off On] to enable or disable the tracking generator.
·[Mode Off On]:
Press
【
Measure
】
→[More 1 of 2]→[Generator]→[Mode CW Track]
to enable or disable the frequency sweep mode of tracking generator.
·[Power 0.0 dBm]:
Press
【
Measure
】
→[More 1 of 2]→[Generator]→[Power 0.0
dBm]; level range -40 dBm~0 dBm, default level 0 dBm, and 1 dB step; set the fixed or
sweep output power of the signal of tracking generator.
·[CW Freq 1.000 GHz]:
Press
【
Measure
】
→[More 1 of 2]→[Generator]→[CW Freq
1.000 GHz]; frequency range 100 kHz~20 GHz, default frequency 1GHz, and 1, 2 or 5
step; valid when the tracking mode is Off. It is the output frequency of the independent
dot frequency generator.
·[Trans Meas]:
Press
【
Measure
】
→[More 1 of 2]→[Generator]→[Trans Meas] to open
the submenu of [Trans Meas] . Refer to the description of “Transmission Measurement”
menu for details.
·[Power Offset 0.0 dB]:
Press
【
Measure
】
→[More 1 of 2]→[Generator]→[Power
Offset 0.0 dB]; range-200 dB~200 dB, default 0 dB, and 1 dB step; valid when tracking
mode is On. Assign a certain offset to the output power of the tracking generator when
gains or losses occur between the generator output and the external device in order to
display the actual power value. This parameter only changes the power readout of the
tracking generator, rather than the actual output of the generator.
·[Frequency
Offset
0.000Hz]:
Press
【
Measure
】
→[More
1
of
2]→[Generator]→[Frequency Offset 0.000 Hz]; range -300 MHz~300 MHz, default 0
Hz and step invalid; valid when tracking mode is enabled; set the frequency deviation
of the output signal of the tracking generator from the current sweep frequency; the
frequency of the tracking generator must be within 100 kHz~20 GHz.
Normalize
Off
On
Relative Ref
0.0dB
Ref Position
0
Scale/Div
10.0dB
Store
Mem Trace
Mem Trace
Off
On
< Back
Trans Meas
·[Normalize Off On]:
Press
【
Measure
】
→[More 1 of 2]→[Generator]→[Trans
Meas]→[Normalize Off On]; range on or off; valid when the tracking mode is enabled
When normalization is enabled, the reference trace will be saved automatically after the
current sweep completes if no reference trace is saved before. In the saving process, a
corresponding message will be prompted on the screen. When normalization is enabled,
the corresponding value of the reference trace will be subtracted from the trace data
after every sweep.
·[Relative Ref 0.0 dB]:
【
Measure
】
→[More 1 of 2]→[Generator]→[Trans
Meas]→[Relative Ref 0.0 dB]; range -200 dB~200 dB, default 0 dB, and 1 dB step;
valid when both the tracking mode and the normalization are enabled. Adjust the
vertical position of the trace on the screen by adjusting the reference level when
normalization is enabled. Different from the reference level in the Amplitude/Scale
menu in the analysis mode, modification of this parameter will not affect the reference
level of the spectrum analyzer.
·[Ref Position 0]:
【
Measure
】
→[More 1 of 2]→[Generator]→[Trans Meas]→[Ref
Position 0]; range 0~10, default 5 and 1 step; valid when both the tracking mode and
the normalization are enabled. Adjust the vertical position of the normalization
reference level on the screen by adjusting the reference position when normalization is
enabled. When it is set to 0, the reference position is at the top; when it is set to 10, the
reference potions is at the bottom; when it is set to 5, the reference position is in the
middle.
·[Scale/Div 10.0 dB]:
【
Measure
】
→[More 1 of 2]→[Generator]→[Trans
Meas]→[Scale/Div 10.0 dB]; range 0.1~20 dB, default 10 dB, 0.1 dB step when less
Содержание 4041 Series
Страница 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 20: ...8 Article I Instructions...
Страница 21: ...9...
Страница 33: ...Chapter III Basic Operations 21...
Страница 96: ...Chapter V Interference Analyzer Mode Option 84 Fig 5 2 RSSI Test Diagram of Interference Analyzer...
Страница 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Страница 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Страница 142: ...130 Article II Technical Instructions...
Страница 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Страница 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Страница 185: ...173 Article III Maintenance Instructions...
Страница 186: ...174...