
Chapter XI Performance Characteristics Test
165
Fig 11-15 Reference Level Uncertainty and Scale Fidelity Testing Setup
1)
Connect the power meter ML2437A and the power probe MA2445D and zero them, and set the calibration
factor frequency to 50 MHz.
2)
Connect the test device according to Fig 15. The spectrum analyzer provides the reference frequency for the
synthesized signal generator 1464C, and the output of the signal generator is connected to the spectrum
analyzer and the power probe respectively through the power divider.
3)
Set the
【
Ampt
】
[Ref Level] on 4041 series, and use the knob or step key
【
↑
】
【
↓
】
. If the maximum
reference level can be set to +30 dBm and the minimum reference level can be set to -120 dBm, the
specifications of the reference level range meet the requirements. In this case, mark
“
√
”
in the
4041 Series
Handheld Spectrum Analyzer Record
, otherwise, mark
“
×
”
in the test result column.
4)
Set the output frequency of 1464C to 50 MHz, and adjust the output frequency to make the power meter
reading be -3 dBm
0.05 dB.
5)
Set the spectrum analyzer as follows:
Center frequency .......................................................................................................................................... 50 MHz
Reference level .............................................................................................................................................. 0 dBm
Span ................................................................................................................................................................ 1 kHz
Logarithmic scale dB/div .................................................................................................................................. 1 dB
Resolution bandwidth ..................................................................................................................................... 10 Hz
6)
Adjust the output amplitude of 1464C to make the signal 2-3 divisions below the reference level of the
spectrum analyzer.
7)
Press
【
Peak
】
【
Marker
】
[Detla] on the spectrum analyzer.
8)
Press
【
↓
】
key on 1464C once.
9)
Set the reference level of the spectrum analyzer to the reference level listed in the
4041 Series Spectrum
Analyzer Record Sheet
; wait for completion of sweep, and press
【
Peak
】
.
BNC cable
Synthesized signal
generator
Spectrum
Analyzer
Adapter
10 MHz output
10MHz IN
2.4 mm cable
Power meter
Sensor
Power divider
Содержание 4041 Series
Страница 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 20: ...8 Article I Instructions...
Страница 21: ...9...
Страница 33: ...Chapter III Basic Operations 21...
Страница 96: ...Chapter V Interference Analyzer Mode Option 84 Fig 5 2 RSSI Test Diagram of Interference Analyzer...
Страница 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Страница 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Страница 142: ...130 Article II Technical Instructions...
Страница 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Страница 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Страница 185: ...173 Article III Maintenance Instructions...
Страница 186: ...174...