Chapter IX Field Strength Measurement Mode (Option)
119
1)
Press
【
Measure
】
→ [FScan] to enable the frequency sweeping mode.
2)
Press
【
Freq
】
→ [Stt Freq] and set the starting frequency of sweeping.
3)
Press
【
Freq
】
→ [Step Freq] and set the step frequency of sweeping.
4)
Press
【
Freq
】
→ [Points] and set the number of sweeping points.
5)
Press
【
BW
】
→
【
BW
】
and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz,
15kHz, 30kHz, 50kHz, 120kHz or 150kHz.
6)
Press
【
Swp/Ant
】
→ [Recall Antenna], and select the antenna factor file.
7)
Press
【
Swp/Ant
】
→ [Dwell Time +∞ Man], and set the dwell time.
8)
Press
【
Swp/Ant
】
→ [Dwell Time Off On] to enable or disable the dwell time.
9)
Press
【
Marker
】
→ [Marker Off On] to enable or disable the marker.
10)
Press
【
Peak
】
to directly set the marker at the maximum point
The dwell time refers to the time of dwelling at the frequency point exceeding the set
limit in the ON state.
The schematic diagram of FScan mode is shown in the Fig below (example).
Fig. 9-2 Schematic Diagram of FScan Mode
9.1.3
MScan
Main operation steps of MScan:
1)
Press
【
List Edit
】
→ [Edit List] and edit the current list.
2)
Press
【
Measure
】
→[MScan] to enable the MScan mode.
3)
Press
【
List Edit
】
→ [Edit List]→
【
Freq
】
and set the step frequency of sweeping.
4)
Press
【
List Edit
】
→ [Edit List]→ [BandWidth] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz,
2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.
5)
Press
【
Swp/Ant
】
→ [Recall Antenna], and select the antenna factor file.
CAUTION
Содержание 4041 Series
Страница 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 20: ...8 Article I Instructions...
Страница 21: ...9...
Страница 33: ...Chapter III Basic Operations 21...
Страница 96: ...Chapter V Interference Analyzer Mode Option 84 Fig 5 2 RSSI Test Diagram of Interference Analyzer...
Страница 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Страница 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Страница 142: ...130 Article II Technical Instructions...
Страница 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Страница 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Страница 185: ...173 Article III Maintenance Instructions...
Страница 186: ...174...