Chapter IV Spectrum Analyzer Mode
40
e)
Set the attenuator to 0 dB.
Press
【
Ampt
】
, [Atten Auto Man], 0 [dB].
Press
【
Peak
】
, [Peak Track Off On].
Press
【
Marker
】
, [Detla].
Fig 4-12 Observation of harmonic distortion
f)
Increase the attenuator setting to 10 dB;
Press
【
Ampt
】
, [Atten Auto Man], 10 [dB].
Note to observe the reading of differential marker, as shown in Fig 4-13. This reading is the distortion difference
generated by the attenuator set at 0 dB and 10 dB. When the attenuator setting is changed, if the difference
frequency marker reading is 1 dB or greater, it means the spectrum analyzer has a certain degree of distortion.
When the differential marker reading is not obvious, the attenuation can be increased.
Содержание 4041 Series
Страница 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 20: ...8 Article I Instructions...
Страница 21: ...9...
Страница 33: ...Chapter III Basic Operations 21...
Страница 96: ...Chapter V Interference Analyzer Mode Option 84 Fig 5 2 RSSI Test Diagram of Interference Analyzer...
Страница 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Страница 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Страница 142: ...130 Article II Technical Instructions...
Страница 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Страница 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Страница 185: ...173 Article III Maintenance Instructions...
Страница 186: ...174...