Chapter IV Spectrum Analyzer Mode
49
Press
【
Peak
】
,
【
Marker
】
, [Detla],
【
Peak
】
, [Next Pk].
The amplitude difference between the main lobe and sidelobe is the sidelobe ratio, as shown in Fig 4-22.
Fig 4-22 Sidelobe ratio displayed beside the marker
e)
Measure the pulse width, which equals the reciprocal of frequency difference between the two sidelobe
envelope peaks:
Press
【
Marker
】
, [Detla],
【
Peak
】
, [Next Pk Right], [Next Pk Right].
At this time, the reciprocal of frequency difference of the read differential marker is the pulse width, as shown in
Fig 4-23. In order to obtain the most accurate pulse width, adjust the marker position manually to measure the
distance between zero crossing points of two adjacent sidelobes. If the resolution bandwidth is reduced to make
the zero crossing point more sharp, the measurement precision will be higher.
Содержание 4041 Series
Страница 1: ...4041 Series Spectrum Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 20: ...8 Article I Instructions...
Страница 21: ...9...
Страница 33: ...Chapter III Basic Operations 21...
Страница 96: ...Chapter V Interference Analyzer Mode Option 84 Fig 5 2 RSSI Test Diagram of Interference Analyzer...
Страница 125: ...Chapter VIII Channel Scanner Mode Option 113 Fig 8 3 Schematic Diagram of List sweep...
Страница 141: ...Chapter IX Field Strength Measurement Mode Option 129...
Страница 142: ...130 Article II Technical Instructions...
Страница 143: ...Chapter IX Field Strength Measurement Mode Option 131...
Страница 184: ...Chapter XI Performance Characteristics Test 172 minimum frequency of 1 Hz and 1 3 steps...
Страница 185: ...173 Article III Maintenance Instructions...
Страница 186: ...174...