Aridus II™ Operator’s Manual
36
5
Disconnect the NEBULIZER SUPPLY line from the quick-connect fitting on the
back of the Aridus II
™
as in (4).
6
Disconnect the SAMPLE OUT Teflon
®
-lined PVC tubing from the ¼ inch
threaded fitting on the back of the Aridus II
™
.
7
If in use, power off and disconnect the ASX–112FR autosampler. Remove the
ASX-112FR from the top of the Aridus II
™
.
8
Remove all screws from the lid of the Aridus II
™
. There are a total of 12 screws
(four top, four left-side, four right-side).
9
Carefully remove the lid of the Aridus II
™
by lifting the lid straight up and off.
10
The membrane desolvator is a modular assembly located in the rear half of the
Aridus II
™
. See Figure 6-4. The membrane desolvator may be completely
removed from the Aridus II
™
and conveniently cleaned on a work bench.
Figure 6-4
. Aridus II
™
– Top View.
11
Completely loosen the three spring-loaded screws which attach the membrane
desolvator to the spray chamber oven. See Figure 6-5.
Membrane
Desolvator
Assembly
Содержание Aridus II
Страница 1: ...CETAC Aridus II Desolvating Nebulizer System Manual Part Number 480144 rev2 Operator s Manual...
Страница 12: ...Aridus II Operator s Manual 12 This page is intentionally blank...
Страница 22: ...Aridus II Operator s Manual 22 Figure 3 11 Aridus II with ASX 112FR Autosampler and QuickWash Accessory...
Страница 26: ...Aridus II Operator s Manual 26 This page is intentionally blank...
Страница 42: ...Aridus II Operator s Manual 42 This page is intentionally blank...
Страница 50: ...Aridus II Operator s Manual 50 This page is intentionally blank...
Страница 60: ...Aridus II Operator s Manual 60 This page is intentionally blank...
Страница 62: ...Manual Part Number 480144 rev2 Printed in USA...