Aridus II™ Operator’s Manual
Chapter 3: Installing the Aridus II™
17
Connecting the Aridus II
™
Drain Tubing
There is one liquid drain line to install on the back of the Aridus II
™
unit. To set
up this drain line:
1
Obtain the peri-pump tubing kit from the Aridus II
™
completion kit.
2
One bag inside the peri-pump tubing kit contains pre-cut tubing attached to
gray-gray color-coded peristaltic tubing. Remove this tubing from the bag.
3
One of the tubing assemblies contains a tan-colored PEEK nut. Install the
PEEK nut into the threaded port labeled DRAIN on the back of the Aridus II
™
unit. Tighten the PEEK nut so it is finger-tight.
4
Install the gray-gray color-coded peristaltic tubing (already attached to the
PEEK nut via a length of Teflon
®
tubing) to the peristaltic pump mounted on
the back of the Aridus II
™
unit. Note that liquid from the drain port should flow
to the bottom of the peristaltic pump and out the top to waste. This is shown
by the arrows next to the pump.
5
The outlet of the drain line may be placed in a suitable waste receptacle (such
as a beaker).
6
Close the shoe of the peristaltic pump and adjust the tension against the pump
rotor head with the spring-loaded screw.
Connecting the ASX-112FR Autosampler to the
Power Source
To connect the ASX-112FR autosampler to a power source:
A voltage-specific external desktop power supply is supplied with each ASX-
112FR autosampler.
To connect the autosampler to a power source, plug the external desktop
power supply cord into the power connector located on the back panel of the
autosampler. Then, plug the power supply’s power cord into the power supply
and into a 100–240-VAC ±10%, 50/60-Hz utility power outlet.
It is important to use the appropriate power cord for your country. Contact
CETAC Technologies if you need a different power cord.
FIRE AND SHOCK HAZARD
Use only the provided desktop power supply. The power supply must be
plugged into an outlet which has a protective ground connection.
WARNING
Содержание Aridus II
Страница 1: ...CETAC Aridus II Desolvating Nebulizer System Manual Part Number 480144 rev2 Operator s Manual...
Страница 12: ...Aridus II Operator s Manual 12 This page is intentionally blank...
Страница 22: ...Aridus II Operator s Manual 22 Figure 3 11 Aridus II with ASX 112FR Autosampler and QuickWash Accessory...
Страница 26: ...Aridus II Operator s Manual 26 This page is intentionally blank...
Страница 42: ...Aridus II Operator s Manual 42 This page is intentionally blank...
Страница 50: ...Aridus II Operator s Manual 50 This page is intentionally blank...
Страница 60: ...Aridus II Operator s Manual 60 This page is intentionally blank...
Страница 62: ...Manual Part Number 480144 rev2 Printed in USA...