Aridus II™ Operator’s Manual
38
Figure 6-7.
Removal of ARGON and NITROGEN gas lines from solenoid valves.
14
Detach the connection to the rear-mounted cooling fan. See Figure 6-8.
Figure 6-8.
Detachment of Cooling Fan Connection.
15
Detach the multi-wire connector to the membrane desolvator. See Figure 6-9.
The membrane desolvator modular assembly may now be completely removed
from the Aridus II
™
cabinet.