Aridus II™ Operator’s Manual
16
Figure 3-3b.
Attachment of Sweep Gas Out Exhaust Line to Trap Bottle.
5
A minimum argon and nitrogen supply of 350 kPa (50 psi, 3.5 bar) is required
to operate the Aridus II
™
. The argon may be teed in from the ICP-MS
instrument argon supply using the ¼
"
union tee. Connect the ¼
"
O.D. tubing,
labeled ARGON and NITROGEN, to the respective supply ports on the back of
the Aridus II
™
. The ¼
"
O.D. tubing for each is simply inserted into the
corresponding press fit connectors. Simply push the tubing into its respective
connector until it is fully secure. Valves are supplied to turn off these gas
supplies when the Aridus II
™
is not in use (See Figure 3-4).
Figure 3-4.
Attachment of Argon and Nitrogen Supply Lines.