44
5.13 Warm-up time
Warm-up time
30 seconds
5.14 Temperature influence
Operation range:
-10 to +50 °C
Maximum error (-10 to +50°C):
0.5 dB
Storage without batteries
-20 to +60 °C
5.15 Humidity influence
Operation range:
30 to 90 %
Maximum error at
30%< R.H.<90% at 40°C and 1 kHz:
0.5 dB
Storage without batteries:
< 93 %
5.16 Magnetic field influence
Influence of the Magnetic Fields
In a magnetic field of 80 A/m (1
oersted) at 50 Hz, the reading
will be under 25 dB(A)
Sound pressure level at which the SC-30
meets
the
requirements
of
radiated
electromagnetic field.
74 dBA
Set of accessories tested in the verification of
the electromagnetic compatibility requirements.
The ones described in chapter
5.24
Configuration for the normal mode of working.
Sound level meter mode, L
AF
function.
Configuration for the reference orientation.
The SC-30 in vertical position,
with the SC-30 main axis
(preamplifier) perpendicular to
the field propagation direction
5.17 Vibration influence
For frequencies from 20 to 1,000 Hz and 1 m/s²:
< 75 dB(A)