Octave band nominal central frequencies
Nominal central frequency
Exact base 10 frequency
31.5 Hz
31.623
Hz
63 Hz
63.096
Hz
125 Hz
125.89
Hz
250 Hz
251.19
Hz
500 Hz
501.19
Hz
1 kHz
1,000
Hz
2 kHz
1,995.3
Hz
4 kHz
2,511.9
Hz
8 kHz
7,943.3
Hz
16 kHz
15,849
Hz
5.8 Measurement range (octave band spectrum analyser)
L
T
Function
for C-130 + PA-13:
Measurement range (with linearity error lower than 0.4 dB):
Octave bands with nominal central frequency
Lower than 63 Hz:
Upper limit:
Lower limit:
From 63 Hz to 8 kHz:
Upper limit:
Lower limit:
Higher than 8 kHz:
Upper limit:
Lower limit:
The noise (elec thermic of the microphone) is, as a minimum,
10 dB lower than the lower limit of the measurement range.
137 dB
31 dB
137 dB
20 dB
137 dB
25 dB
for C-250 + PA-14:
Measurement range (with linearity error lower than 0.4 dB):
Octave bands with nominal central frequency
Lower than 125 Hz:
Upper limit:
Lower limit:
From 125 Hz to 8 kHz: Upper limit:
Lower limit:
Higher than 8 kHz:
Upper limit:
Lower limit:
The noise (elec thermic of the microphone) is, as a minimum,
10 dB lower than the lower limit of the measurement range.
137 dB
36 dB
137 dB
23 dB
137 dB
36 dB