SC-30
User’s Manual
41
5.9 Microphone
C-130
½
″
condenser microphone
Polarisation
200 V
Nominal capacitance
22.5 pF
Nominal sensitivity:
17.5 mV/Pa
±
0.5dB in reference
conditions
Effect of the windscreen
< 1 dB for frequencies < 10 kHz
< 3 dB for frequencies < 12.5 kHz
Constant pressure to free field correction
Frequency (Hz)
Correction (dB)
Frequency (Hz)
Correction (dB)
31.5
-0.1
2,000
0.3
63
-0.1
4,000
1.2
125
0.0
8,000
3.2
250
0.0
12,500
6.3
500
0.0
16,000
8.8
1,000
0.1
C-250
½
″
prepolarised condenser microphone
Polarisation
0 V
Nominal capacitance:
17.0 pF
Nominal sensitivity:
46.4 mV/Pa in reference conditions
Effect of the windscreen
< 1 dB for frequencies < 10 kHz
< 3 dB for frequencies < 12.5 kHz
Constant pressure to free field correction
Frequency (Hz)
Correction (dB)
Frequency (Hz)
Correction (dB)
125
0.0
2,000
0.4
250
0.0
4,000
1.2
500
0.0
8,000
3.4
1,000
0.1