Instruction Manual CertoClav Vacuum Pro 45 l — Model STE-45-T
55
11.6 Sensor calibration
The sensor calibration should be calibrated by trained personnel
only.
To calibrate the sensors, navigate to the Settings Menu by clicking on the
symbol
at the top right of the screen and pressing on “Calibration”.
Press on the offset you wish to adjust. The offset parameters must be in the interval [-
3°C; 3°C] and [-5 kPa; +5kPa]. Then press on “Apply”.
11.7 GLP and GMP data
To view the GLP and GMP settings, navigate to the Settings Menu by clicking on the
symbol at the top right of the screen and pressing on GLP and GMP data.
11.7.1 Header
The header setting activates the printout of the GLP header. Up to four text lines are
available. If this setting is activated, the header will be displayed at the top of each
sterilisation protocol and printed out.
11.7.2 Autoclave designation
Designate the autoclave in this setting. An alphanumeric text can be entered. The
standard name is empty.
11.7.3 Project designation
Designate your project in this setting. An alphanumeric text of up to 25 characters can
be entered. The standard designation is empty.