11.3 Sterilisation settings
11.3.1 Customising sterilisation programs
To customise a sterilisation program, navigate to the Settings Menu by clicking on the
symbol at the top right of the screen. Press on “Sterilization” and then “Edit user
defined program”.
Now define the program parameters individually corresponds to your requirements:
•
Number of prevacuum pumps between 1 and 10
•
Sterilisation temperature between 105 °C and 134 °C
•
Sterilisation time between 4 and 60 minutes
•
Drying time between 1 and 25 minutes
Then press on “Apply” to save the program parameters or on “Cancel” to return to the
Settings Menu and cancel the process.
11.3.2 Material tests and automatic repetitions