Manual_CLASSIC_DE_EN_V4
Page 23 of 28
CertoClav Sterilizer GmbH
Peintner Strasse 10
A-4060 Leonding / Austria
Operation
Before first use
READ INSTRUCTIONS BEFORE FIRST USE!
Run the first cycle without load; with water only.
Loading the autoclave
•
Fill chamber with 1 - 1.5 liters of distilled water.
•
It is possible to use more water, this will however result in an increased heating time and cooling down
period.
•
In order to avoid calcification of your device, we recommend the use of distilled water and
decalcification on a regular basis.
•
Place the instrument tray inside the chamber and put the load on top. It is important that steam can
circulate freely, therefore, it is necessary to provide enough space between the load and the chamber
wall.
Attention: Do not autoclave tightly sealed containers (especially glass bottles) because there is no
pressure compensation.
Closing and locking
•
Check if the gasket is inserted in the lid.
•
Press the red knob and turn it clockwise to
“
OPEN
”
.