Autoclave
–
safety features
Secure locking mechanism and steam release valve inside the handle
To prevent opening under pressure as well as pressure build up when the lid is not properly closed.
Turning the red knob to “OPEN” automatically opens the steam valve and
steam is being released. As long
as the unit is not properly and securely closed, the steam release valve remains open, preventing any
pressure build-up.
Pressure control valve:
In case of exceedance of the highest working pressure of 2,7 bar by more than 10%, pressure is released
through this safety mechanism.
Safety valve:
In the event that the pressure control valve fails, the safety valve opens at 3,6
–
4 bar to ensure the safety
of the user. This safety valve closes automatically as soon as operating pressure has been reached. If this
happens the pressure control valve needs to be changed immediately, as it is defect.
If this occurs DO NOT use the device and return it to the dealer or the manufacturer for
examination and a security check. It is dangerous to work with such a unit.
Technical data
Technical data
Material
Stainless steel 316L
Certificates
CE
Dimensions
Volume
18 liters
Height
60,5 cm
Chamber diameter
26cm
Width
44,5 cm
Chamber height
35cm
Depth
30 cm
Useable height chamber
30 cm
Weight
8,8 kg
Working data
Max. working temp
140°C
Current
230V, 50/60 Hz
Max. working pressure
2,7 Bar
Power
1300 Watt
Test pressure
4,1 Bar
Operation time
S2 30 min
Directives and standards
Tested according:
•
2014/30/EU
–
EMC
–
Associated stan
dards:
o
EN 61326-1:2013 Electrical Equipment for measurements, control and laboratory use - EMC-
Requirements Part 1: General requirements
o
EN 61000-3-2:2006 + A1:2009 + A2:2009 Electromagnetic Compatibility (EMC)
—
Part 3- 2: Limits -
Limits for harmonic current emissions (Equipment input current equal to or less than <= 16A per
phase)
o
EN 61000-3-3:2013 Electromagnetic Compatibility
—
Part 3- 3: Limits
—
Limits for voltage
fluctuation and flicker on the public low-voltage system (equipment having an input current to or
less <= 16 A per phase), and not subject to conditional connection.
•
EN 61010-1:2010
•
RoHS 2011/65/EU: Directive on the restriction of the use of certain hazardous substances