Cerlic CBX Скачать руководство пользователя страница 8

fghj 

 

 
 
 
 
 
 

   

8   

   

      

CBX   

   

   

   

Rake   interval   

The   rake   limit   switch   starts   a   measuring   cycle.   When   the   rake   is   passing   the   rake   limit   
switch   the   CBX   will   execute   a   sampling   cycle.      It   is   possible   to   configure   the   number   of   
pulses   from   the   rake   switch   that   is   needs   for   a   measuring   cycle   to   start.   This   parameter   

is   called   “

trig   int

”.   It   can   be   used   for   delaying   the   time   between   the   samples,   normally   

you   will   take   a   sample   every   30   -­‐   45   minute.   The   pulse   from   the   rake   limit   must   not   be   

longer   than   one   complete   measuring   cycle   on   the   CBX.   

Emergency   retraction   of   the   cable      

If   the   CBX   for   some   reason   does   not   retract   the   cable,   then   it   is   possible   to   do   manually.   

But   always   try   to   switch   off   the   main   power   to   the   CBX   for   10   seconds   and   then   back   on   
line   again.   The   CBX   will   always   start   with   retracting   the   cable   to   “

home   position

”   after   a   

power   loss.   If   the   cable   still   does   not   retract   use   the   emergency   motor   control   switch   

that   is   located   on   the   box   for   the   terminal   block   inside   the   CBX.   The   switch   can   be   
located   on   top   of   the   terminal   box   or   under   the   lid.   As   a   safety   feature,   you   must   press   
and   hold   the   door   switch   in   during   manual   operation.   

   

NOTE!   

As   a   safety   you   must   press   and   hold   the   door   switch   in   during   manual   operation.   
   

Configuration   terms   

Home   position,   A   

The   sensor   will   always   return   to   this   position.   

Blind   zone,   B   

Sludge   or   foam   detected   in   this   zone   during   lowering   of   the   sensor   is   not   registered.   
Unwanted   interference   is   prevented   by   setting   this   zone   to   be   1’   (0.3m)   below   the   

normal   liquid   surface.   Sensor   does   not   start   to   measure   until   it   reaches   this   distance.   

Max   depth,   C   

This   is   the   maximum   distance   from   the   home   position.   The   sensor   will   stop   at   this   depth   

if   no   sludge   is   detected.   The   sensor   might   stop   and   return   at   a   shorter   distance   if   a   rake   
guard   offset   is   entered   (F).   

Sludge   level,   D   

The   sludge   level   distance   from   bottom   found   when   the   preset   sludge   concentration   is   
reached.   

Measured   range,   E   

The   output   signal   will   vary   from   4-­‐20   mA   within   this   range.   

Raker,   F   

This   denotes   the   safe   height   from   the   bottom   (Max   depth)   to   a   point   where   the   sensor   
always   must   stop   and   then   turn   back   to   home   position   to   alleviate   hitting   the   bottom   
rake.   

Fluff,   G   

The   fluff   layer   or   depth   found   when   the   present   fluff   concentration   is   reached.   
   

   

Содержание CBX

Страница 1: ...2015 05 28 C81D5EN16 CBX Cerlic CBX Sludge Blanket Monitor ...

Страница 2: ...RFACE 6 Configurations 6 LED INDICATION 7 OPERATION 7 CBX MODE 7 Rake interval 8 EMERGENCY RETRACTION OF THE CABLE 8 CONFIGURATION TERMS 8 Home position A 8 Blind zone B 8 Max depth C 8 Sludge level D 8 Measured range E 8 Raker F 8 Fluff G 8 BB1 BB2 MENU FOR CBX 9 Settings 9 CBX Setup 10 Calibration 11 Cleaning 11 Scale Alarm 11 System 11 Timer 12 INFORMATION DISPLAY 12 PROFILE 12 CALIBRATION 14 C...

Страница 3: ...CHNICAL SPECIFICATION FOR 4 20MA MODULE 19 MAINTENANCE 19 MANUAL OPERATION OF CBX 19 Maintenance schedule 19 Cable and sensor 19 TROUBLESHOOTING 20 ALARM 20 APPENDIX 1 MOUNTING OF CBX SUPPORT STAND 20 APPENDIX 2 MOUNTING OF FLUSH PIPE 21 APPENDIX 3 WIRING DIAGRAM 21 APPENDIX 4 SAFETY INSTRUCTIONS 22 CONTACT INFORMATION 23 WARRANTY 23 SPECIFICATION CBX 23 DIMENSIONS IN MM 24 APPENDIX 7 SERVICE FORM...

Страница 4: ...easure sludge and fluff blanket depth in clarifiers thickeners etc in water and wastewater plants as well in other facilities It is possible to display measured values as text or as a graphic image of the sludge profile Working principle The sensor is sent down into the clarifier thickener and measures the suspended solids as it descends and sends the data via a communication cable to the BB2 cont...

Страница 5: ...32 M12 Cable 1 5m 5 20805752 M12 Cable 10m 33 20805510 M12 Cable 30m 100 20850727 Sensor with 33 cable 11305888 Flushing Valve 115V 11705516B Motor 115V 21750924 Fan 115V 21750926 Door key 20305403 Accessories Cerlic Sensor Cleaning Fluid CSC 10905944 CBX Jacket for cold weather 11205969 Starting up Connect the CBX to the BB1 BB2 and connect 110 1 60 power to both units BB1 BB2 must have program v...

Страница 6: ...t with the sensor or cable Cerlic Controls AB does not assume any responsibility for damages caused by the absence or malfunction of the rake guard limit switch User interface The BB1 BB2 will show the CBX as a standard X sensor on the display and it will show both sludge and fluff level in text on the display Configurations The figure below shows the different terms used for distances that are in...

Страница 7: ...figurations are done via BB1 BB2 that is connected to the CBX All settings shall be done and saved done by exiting the menu before a new sampling is preformed Depth and concentration is shown on the BB1 BB2 on line during a sampling cycle CBX Mode There is two different modes the CBX can be set to Setup and Trig Trig is the normal working mode In this mode the CBX will take a sample on every exter...

Страница 8: ...l box or under the lid As a safety feature you must press and hold the door switch in during manual operation NOTE As a safety you must press and hold the door switch in during manual operation Configuration terms Home position A The sensor will always return to this position Blind zone B Sludge or foam detected in this zone during lowering of the sensor is not registered Unwanted interference is ...

Страница 9: ... Time s Integration time or dampening can be set up to 999 seconds This is integration in BB1 BB2 The value coming from CBX is normally also an average Manual up down Enables the BB1 BB2 to control the sensor in the CBX to go UP and DOWN In this mode the depth and concentration are displayed in the BB1 BB2 Must change MODE to TRIG to activate Depth Unit Unit used for distances in CBX settings in m...

Страница 10: ...sludge is detected Distance C This travel distance will be shortened if a raker offset exists distance F Blind zone Distance B Distance to where sensor must travel before it starts to measure solids Sludge or foam detected in this zone during lowering is not registered Unwanted false blanket depth is prevented by setting this zone to be approximately 25cm 1 ft below liquid surface Rake height Dist...

Страница 11: ...lids value for Sample Cleaning Cleaning Yes No Yes flush cleaning is switched ON No the flush is switched OFF Manual up down Enables the BB1 BB2 to control the sensor in the CBX to go UP and DOWN In this mode the sensor is switched off and can be left out for an unlimited time for cleaning Must change MODE to TRIG to activate Scale Alarm Max Max level or 20 mA output distance E in fig 2 Min Min le...

Страница 12: ...st sampling available on software on the BB1 BB2 3 23 and later The profile shows fluff and sludge level the top number is fluff level and the lower number is sludge level measured from the tank bottom and up fig 3 Profile CBX users can get the sludge profile using the 4 output signals on the BB2 It is updated at the end of each sampling cycle channel 3 4 are on line during sampling This option re...

Страница 13: ...anket signal updated every time a new measurement is done scale alarms Max depth Channel 3 Concentration on line during a sample otherwise 0 Scale is locked to 0 30000 mg l Channel 4 Depth of sensor on line during sampling otherwise 20mA Scale is same as scale alarms Max depth See graph below how a typical profile can look like when you have about 1ft of sludge and 3ft fluff on top of the sludge S...

Страница 14: ... to Take sample in the calibration menu and choose either Zero for clean water or sample depending on type of calibration to be preformed The CBX will collect samples and the BB1 BB2 shows this with a pop up window 5 In the case of sludge calibration enter the Sample concentration from the lab test in the calibration menu 6 Set the CBX in Trig mode Calibration Modes and back out of the menu to sav...

Страница 15: ... mounted on the cable drum is powered by contact free power inductive from the pickup board It communicates with the main board via the pickup board and it communicates with the sensor through an optical RS 485 interface fig 7 Sensor The sensor is connected to a cable that supplies it with power and transmits the communication signals The sensor contains a near infrared light source and a receiver...

Страница 16: ...ind the drum mounting plate there is a heater which is used in cooperation with the fan to maintain a stable temperature inside the cabinet The fan is mounted on the top right side of the cabinet fig10 fig 10 Flushing tube valve The flushing tube contains a flushing nozzle ring that sprays cleaning water around the sensor and cable during the return or rising of the sensor to the home position The...

Страница 17: ... located on the box for the terminal block inside the CBX The switch can be located on top of the terminal box or under the lid fig 12 fig 12 Appendix 6 Optional 4 20mA output module for BB1 BB2 Introduction The BB1 BB2 4 20mA module is used to expand the BB1 BB2 central unit from two to four 4 20mA loops It is assumed that the user is familiar with the BB1 BB2 and 4 20mA technology Module overvie...

Страница 18: ...tartup display and under Exp module in the BB2 Settings menu fig 14 Mounting the module in the control box Wire connections Connect the 4 20mA loops to the screw terminals according to the table below Pin Name Function No Name Function 1 Ch3 Channel 3 return 2 Ch3 Channel 3 positive 3 Ch4 Channel 4 return 4 Ch4 Channel 4 positive Getting started with the 4 20 mA module A step by step guide to get ...

Страница 19: ...hanical parts need to be inspected at regular intervals and the optical parts may need cleaning Manual operation of CBX It is possible initiate a manual trig of a sample By holding down the down arrow for 5 sec on the BB1 BB2 The CBX starts a sampling cycle ASAP and returns the sensor back home and normal mode If more then one sensor BB2 is connected this function will be disabled ê Starts lowerin...

Страница 20: ...nics to stop then a reset may be performed by switching the power OFF and then ON This can easily be done by opening the CBX door and then closing Alarm A number of errors can be detected and shown on the BB1 BB2 The alarm error is displayed in plain text and the alarm contact is closed It may stop the operations of the CBX and it is reset by pressing ENTER on the BB1 BB2 SENSOR TILT means that th...

Страница 21: ...ose or pipe of at least diameter to fitting The water pressure must not exceed 6 bars 90 psi A last check before starting the CBX 1 Check that the sensor cable is properly wound on drum and cable guide 2 Check that all screws and nuts are properly tightened Appendix 3 wiring diagram Connection box for power supply rake limit switch fig 16 Leveling screws U bolts for 1 handrail or 4 6 angled handra...

Страница 22: ...ried out by authorized personnel Certificate of conformity The CBX conforms to the following EC Directive s when installed in accordance with the installation instructions contained in the product documentation 73 23 EEC Low Voltage Directive as amended by 93 68 EEC 89 336 EEC EMC Directive as amended by 92 31 EEC and 93 68 EEC Standards EN 61000 6 4 2001 Electromagnetic compatibility EMC Part 6 4...

Страница 23: ...is determined that the problem is any fault in manufacturing or equipment during the warranty period Specification CBX Manufacturer Cerlic Control AB Sweden Made in Sweden Name CBX P N 12705845A 220V 1 50 Europe 12705845B 110 V 1 60 US Dimensions See drawing page 16 Weight 21 Kg 46 Ib incl accessories Enclosure NEMA4 IP65 Stainless steel Measuring principle Sounding by optical sensor with straight...

Страница 24: ...pen 24VDC is supplied from CBX Sensor Waterproof to 10m 30 ft in stainless steel 2343 with cable in PUR Motor 1800 rpm synchronous with gearbox 1 100 1500 rpm at 230V Lowering speed Max 15cm s 12 5 cm s at 50 Hz Full cycle time for 10 meters 33 is 3min Measuring range 0 10 m 33 Accuracy sensor 5 Full Scale Accuracy level 0 5 Full Scale Dimensions in mm fig 18 ...

Страница 25: ...__________ Phone ___________________________________________________________ E mail ___________________________________________________________ Product ___________________________________________________________ Serial no ___________________________________________________________ Position Ref ___________________________________________________________ In order to get a fast and accurate service p...

Страница 26: ...__ Fluff Conc _____________________________________________________ Mode _____________________________________________________ Mains Freq _____________________________________________________ Trigger _____________________________________________________ Max depth _____________________________________________________ Blind zone _____________________________________________________ Rake height _____...

Отзывы: