Compass Cast: Software Codes

UF 

Bluetooth Radio Check Failure

SF   

Software Failure

HF 

Hardware Failure

bF 

Battery Failure 

EE 

Environment Error - Device outside operating temperature

CE 

Calibration Error – Pressure values outside operating range

Lb3 

Less than 8 hours battery life remaining

Lb2 

Less than 6 hours battery life remaining

Lb1 

Less than 4 hours battery life remaining

Lb0 

Less than 2 hours battery life remaining

Compass MAP: LED Codes

Solid LED........................................................................  Wireless connection established
Two Flashes Every Four Seconds ...................................  Scanning for wireless connection
Flashing Sixteen Times Per Second ................................  Compass MAP non-functional

Compass CAST FCCID: 2AF4Z-CWHG001
Compass MAP FCCID: 2AF4Z-CWM001

Содержание Compass Cast

Страница 1: ...abled physiological pressure transducer with integrated digital display The Compass Monitor Accessory Plug Compass MAP receives the wireless pressure signal from the Compass Cast and inputs the pressu...

Страница 2: ...our seconds to indicate that it is scanning for a compatible Compass Cast 5 Zero the pressure signal on the patient monitor per the monitor s IFU 6 Remove Compass Cast from its sterile package 7 Remov...

Страница 3: ...it to atmospheric pressure and then cycle device power it off and then power back on to recalibrate To Power Off and Recalibrate Press and hold button for two seconds until is displayed on the LCD Fi...

Страница 4: ...he application attach accessory e g flushing syringe stopcock extension or IV tubing etc to the proximal female luer lock port of the Compass Cast per hospital protocol Figure 3 5 Depending on the app...

Страница 5: ...power off the device To recalibrate depress the power button for approximately one second until the LCD reads 00 7 If procedure warrants insert guidewire up to 0 038 diameter through the Compass Cast...

Страница 6: ...mm displayed will change from flashing to off The device must be power cycled for it to restart advertising for a wireless connection Once a Compass MAP has established a wireless connection to a Com...

Страница 7: ...104 F Storage Temperature 29 to 49 C 20 120 F Light Sensitivity 1 mm Hg Lithium Ion Battery Voltage 3 Volts Display symbol for positive pressure symbol for negative pressure Bar 1 1 mm Hg Bar 12 12 m...

Страница 8: ...or Pressure values outside operating range Lb3 Less than 8 hours battery life remaining Lb2 Less than 6 hours battery life remaining Lb1 Less than 4 hours battery life remaining Lb0 Less than 2 hours...

Отзывы: