Suggested Instructions for Use:

Compass Cast Use without Compass MAP 

1.  Remove Compass Cast from its sterile package.

2.  Remove the proximal luer cap 

(Figure 1)

.

3.   Activate and calibrate Compass Cast by depressing the 

power button for approximately one second until the LCD 

reads “00” 

(Figure 2)

.

WARNING: Activate and auto-calibrate Compass Cast 

prior to attaching accessories to avoid incorrect  

baseline pressure.

4.   Depending on the application, attach accessory  

(e.g. flushing syringe, stopcock, extension or IV tubing, 

etc.) to the proximal, female luer lock port of the  

Compass Cast per hospital protocol 

(Figure 3)

.

5.   Depending on the application, connect the distal, male luer 

lock end of the Compass Cast to accessory (e.g. needle, 

distal lumen of a central line, extension set or needleless 

valve in contact with patient) 

(Figure 4)

.

WARNING: Pressure reading is NOT physiological if either 

the distal or proximal luer connections of the device is 

open to atmospheric pressure.

2

1

00

mm

00

mm

4

3

Compass CAST FCCID: 2AF4Z-CWHG001
Compass MAP FCCID: 2AF4Z-CWM001

Содержание Compass Cast

Страница 1: ...abled physiological pressure transducer with integrated digital display The Compass Monitor Accessory Plug Compass MAP receives the wireless pressure signal from the Compass Cast and inputs the pressu...

Страница 2: ...our seconds to indicate that it is scanning for a compatible Compass Cast 5 Zero the pressure signal on the patient monitor per the monitor s IFU 6 Remove Compass Cast from its sterile package 7 Remov...

Страница 3: ...it to atmospheric pressure and then cycle device power it off and then power back on to recalibrate To Power Off and Recalibrate Press and hold button for two seconds until is displayed on the LCD Fi...

Страница 4: ...he application attach accessory e g flushing syringe stopcock extension or IV tubing etc to the proximal female luer lock port of the Compass Cast per hospital protocol Figure 3 5 Depending on the app...

Страница 5: ...power off the device To recalibrate depress the power button for approximately one second until the LCD reads 00 7 If procedure warrants insert guidewire up to 0 038 diameter through the Compass Cast...

Страница 6: ...mm displayed will change from flashing to off The device must be power cycled for it to restart advertising for a wireless connection Once a Compass MAP has established a wireless connection to a Com...

Страница 7: ...104 F Storage Temperature 29 to 49 C 20 120 F Light Sensitivity 1 mm Hg Lithium Ion Battery Voltage 3 Volts Display symbol for positive pressure symbol for negative pressure Bar 1 1 mm Hg Bar 12 12 m...

Страница 8: ...or Pressure values outside operating range Lb3 Less than 8 hours battery life remaining Lb2 Less than 6 hours battery life remaining Lb1 Less than 4 hours battery life remaining Lb0 Less than 2 hours...