Centurion Compass Cast Скачать руководство пользователя страница 6

Connectivity Details

When Compass MAP enters scanning mode after being powered on with the connection to a patient 
monitor, it will select the strongest available Compass Cast signal.

Once a Compass Cast and a Compass MAP are paired, the devices will no longer be able to pair with 
other Compass devices unless they are power cycled to force them back into general scanning mode.

If the connection is temporarily lost (e.g. because the Compass MAP is disconnected and therefore 
powered down, and reattached to a different patient monitor), the previously paired devices will 
automatically reinitiate a connection.

If the Compass Cast is used without the Compass MAP, after 10 minutes of broadcasting without  
a paired connection, Compass Cast will no longer advertise and the “mm” displayed will change  
from flashing to off.

If Compass Cast is unable to establish a wireless connection, after 10 minutes of broadcasting,  
Compass Cast will no longer advertise and the “mm” displayed will change from flashing to off.  
The device must be power cycled for it to restart advertising for a wireless connection.

Once a Compass MAP has established a wireless connection to a Compass Cast, it will only pair  
to that particular Compass Cast. The Compass MAP must be power cycled to have it search for  
any available Compass Cast.

Device Power Off and Recalibration

Cleaning Instructions

Press and hold button for two seconds until “---” is displayed on the LCD, then release button  
to power down device. 

The Compass MAP, the data receiver and monitor interface, is non-patient contacting and intended for 
multiple uses.   

1.  Inspect the Compass MAP for signs of damage, wear, or abuse.

2.   Do not use if worn or damaged, e.g. connection pins bent or missing,  

or if the casing seal’s integrity is compromised. 

3.  The Compass MAP may be low-level disinfected per hospital protocol. 

4.  Do not immerse Compass MAP in solutions.

5.  Do not super saturate the cloth or the device. 

6.  Avoid liquid contact with the pin connectors.

7.  Do not use abrasive or scouring materials.

8.  Do not use staining solutions.

Compass CAST FCCID: 2AF4Z-CWHG001
Compass MAP FCCID: 2AF4Z-CWM001

Содержание Compass Cast

Страница 1: ...abled physiological pressure transducer with integrated digital display The Compass Monitor Accessory Plug Compass MAP receives the wireless pressure signal from the Compass Cast and inputs the pressu...

Страница 2: ...our seconds to indicate that it is scanning for a compatible Compass Cast 5 Zero the pressure signal on the patient monitor per the monitor s IFU 6 Remove Compass Cast from its sterile package 7 Remov...

Страница 3: ...it to atmospheric pressure and then cycle device power it off and then power back on to recalibrate To Power Off and Recalibrate Press and hold button for two seconds until is displayed on the LCD Fi...

Страница 4: ...he application attach accessory e g flushing syringe stopcock extension or IV tubing etc to the proximal female luer lock port of the Compass Cast per hospital protocol Figure 3 5 Depending on the app...

Страница 5: ...power off the device To recalibrate depress the power button for approximately one second until the LCD reads 00 7 If procedure warrants insert guidewire up to 0 038 diameter through the Compass Cast...

Страница 6: ...mm displayed will change from flashing to off The device must be power cycled for it to restart advertising for a wireless connection Once a Compass MAP has established a wireless connection to a Com...

Страница 7: ...104 F Storage Temperature 29 to 49 C 20 120 F Light Sensitivity 1 mm Hg Lithium Ion Battery Voltage 3 Volts Display symbol for positive pressure symbol for negative pressure Bar 1 1 mm Hg Bar 12 12 m...

Страница 8: ...or Pressure values outside operating range Lb3 Less than 8 hours battery life remaining Lb2 Less than 6 hours battery life remaining Lb1 Less than 4 hours battery life remaining Lb0 Less than 2 hours...