Suggested Instructions for Use:
Compass Cast Use without Compass MAP (concluded)
6. Pressure is displayed on the Compass Cast LCD
(Figure 5)
.
NOTE: If Compass Cast pressure reading does not return to
zero after removal from accessories, flush device with saline,
open it to atmospheric pressure, and then cycle device (power
it off and then power back on) to recalibrate.
To Power Off and Recalibration
Press and hold button for two seconds until “---” is displayed
on the LCD
(Figure 6)
, then release button to power off
the device. To recalibrate, depress the power button for
approximately one second until the LCD reads “00”.
7. If procedure warrants, insert guidewire (up to 0.038˝
diameter) through the Compass Cast guidewire port
and into vessel. Smaller guidewires may necessitate
the use of the guidewire insertion tool accompanying
the guidewire holder
(Figure 7)
.
WARNINGS
Do not reinfuse blood or other bodily fluid into
the patient after any such fluid is drawn into the
Compass Cast.
Do not reuse Compass Cast for different procedures
with different patients. Use a new Compass Cast
device for each procedure with each patient.