RUN-OUT ADJUSTMENT:

STEP 4:

     Position the dial indicator, as shown in the photo below, with the finger of the dial indicator

on the front and center of the stylus ball. Insert the 3/32” hex wrench in the run-out adjustment set 
screw directly above the finger of the dial indicator.  Loosen the set screw by turning the hex wrench 
counter clockwise while watching the dial indicator reading decrease. Turn the hex wrench clockwise 
and watch the dial indicator reading increase. 

Run-out Adjustment set-up and effect of Set Screw Movement

STEP 5:

Rotate the spindle by hand (without touching the probe) so that the probe spins through a

full 360 degrees and watch the dial indicator to locate the high and low run-out rotation positions of the
stylus. Adjust the dial indicator so that the entire run-out of the tip can be seen on the dial. If the run-
out exceeds the range of the dial then begin at the highest point of the run-out and set the dial indicator 
so it is at the limit of its range at this point.  

STEP 6:

Rotate the probe so that the nearest run-out adjustment set screw is above the finger tip 

of the indicator. Using the 3/32" hex wrench slowly turn the setscrew directly above the indicator 
finger, in the counter clock-wise direction, the dial indicator will show the reduction in the run-out. 
Stop loosening this set screw and tighten the set screw that was nearest the low run-out position. Do not
allow the probe to become loose on the mounting shank.  

STEP 7:

Repeat steps 5 and 6 until no run-out of the tip is visible on the dial indicator needle. If 

the screws become too tight before you have finished the adjustment, 

do not force them

, instead 

loosen the setscrew that corresponds with the highest reading on the dial indicator, (refer to figure 3.2) 
and tighten the set screw at the lowest reading.  Only adjust a set screw when it is directly above the 
indicator finger. This way you can see the full effect of the adjustment you are making. 

4

Содержание DP-4

Страница 1: ...DP 4 Touch Probe Operator s Manual U S Patent 6553682 Rev 2019 05 28 Copyright 2003 2020 Centroid Corporation Howard PA 16841...

Страница 2: ...nfirm machine is responding to DP 4 input by running an automatic cycle without the probe in the spindle and manually touching the DP 4 stylus mount boss Refer to control operator s manual on how to r...

Страница 3: ...Diameter Calibration Page 6 Performance Characteristics Page 8 Maintenance and Care Page 9 Shipping Instructions Page 9 Kit Contents Page 10 Common Cables Page 11 Connection Diagram universal Page 12...

Страница 4: ...orms OUTPUT FEATURES The output signal is a sink type compatible with most control inputs that can be configured for normally open or closed logic See CONNECTION DIAGRAM UNIVERSAL section This choice...

Страница 5: ...2 56 4mm Mounting shank diameter and length D 0 5 12 7mm L 1 47 37 2mm Stylus mount thread M3 Output signal Switch FET open drain 30V 50mA Output logic Selectable normally open or closed Power supply...

Страница 6: ...he DP 4 Probe into the dedicated 0 500 tool holder STEP 2 Install the complete assembly into the milling machine spindle without the retractile cord STEP 3 Install the 2 5 or 5mm stylus in your DP 4 p...

Страница 7: ...then begin at the highest point of the run out and set the dial indicator so it is at the limit of its range at this point STEP 6 Rotate the probe so that the nearest run out adjustment set screw is...

Страница 8: ...nt begin by completing the Dialing in procedures for the probe tip Measure the run out at the base of the stylus stem the end away from the tip with the dial indicator If the run out is greater than t...

Страница 9: ...n the desired direction and finally Cycle Start The control will jog the stylus in the direction of the single axis move selected Touch the stylus once and the control will stop and reverse travel the...

Страница 10: ...appear on the screen that will display the measured diameter of the bore Subtract this number from the known diameter of the gauge bore The resulting number will be the calibrated diameter of the prob...

Страница 11: ...tion procedures then accuracy becomes dependent on repeatability PRE TRAVEL VARIATION RADIAL DIGITIZING PRECISION BORE Typical pre travel variation for DP 4 probe showing variation over 720 degrees of...

Страница 12: ...d in or aligned then the probe should be sent in for evaluation and repair Probing errors may be caused by many factors If the errors are random in nature check surface condition of stylus tip and pro...

Страница 13: ...auge p n 3193 Size and type may vary 5mm Ruby tip stylus M3 thread 40mmlength p n 3360 3 32 hex wrench p n 2898 DP 4 Probe complete kit with protective case and accessories Inset shows outside of case...

Страница 14: ...are needed for new Centroid TM systems See the appropriate appendix in this manual for additional required adapters connection diagrams and parameter settings for your installation 11 Probe Cable P N...

Страница 15: ...as follows 1 Power DC voltage supply positive terminal connected to pin 2 negative to pin 5 2 Signal Current sinking type pin 3 appears connected to pin 5 when signal is active Pin 3 appears unconnect...

Страница 16: ...have a much better feel for where these parameters need to be set to achieve the results to meet your job requirements The X Y and Z stepover The smaller the stepover distance the longer the digitizin...

Страница 17: ...end of the program You can also use the text editor to turn coolant on and off by adding M7 MIST or M8 FLOOD at the beginning of the program and M9 COOLANT OFF at the end Remember you can have only on...

Страница 18: ...n enable 155 0 Probe type enable 186 1 Probe stuck retry disable 366 1 Probe Deceleration Multiplier Probe Jog Parameters control jog speed when probe is connected Probe must decelerate to stop when c...

Страница 19: ...ate 15 3 76 2 Slow Probing Rate 16 5 127 Maximum Search Distance 18 50771 PLC input Probe Detect Spindle Inhibit 120 0 020 0 508 Probe stuck clearance amount 121 0 020 0 508 Grid digitizing minimum Z...

Страница 20: ...unt nominal 14 30 762 Fast Probing Rate 15 3 76 2 Slow Probing Rate 16 5 127 Maximum Search Distance 18 15 PLC input Probe Detect Spindle Inhibit 120 0 020 0 508 Probe stuck clearance amount 121 0 020...

Страница 21: ...ose that require a current source signal The circuit shown below can be used to convert the probe tripped signal from current sink to current source if necessary The resistor value is calculated to pr...

Отзывы: