Center 265 True RMS Leakage Current Clamp Meter 

 

 

 

5.  BUTTON INSTRUCTION 

5-1  HOLD & LPF Function 

It is possible to freeze the displayed value by pressing on the "

HOLD

button. Press the "

HOLD

" button again to exit the Hold mode. 

To  activate  the  Low  Pass  Filter  (LPF)  feature,  press  and  hold  the 

LPF

” button until the symbol displays. To deactivate the LPF feature, 

press and hold the “LPF” button for 2 seconds. The meter will return 
to normal mode of operation. 

 

The Cut-Off frequency of the low pass filter is about 160 Hz with 
attenuation characteristic of approx. -24 dB/octave. 
 

Note:

 The LPF function operates on the AC A and AC mA ranges. 

 

 

5-2  PEAK Hold  &  

 /    Function Selection 

       Press "

PEAK

" button to enter the PEAK mode, the meter records the 

peak maximum value. Press the "

 PEAK

 " button again, to exit the 

peak mode. 

 

When  you  turn  the  rotary  switch  to  the  Ohms  /  Continuity  Test 
position, press the "

PEAK

" button to select Resistance measurement, 

press the "

PEAK

" button again to select continuity test with buzzer. 

 

5-3  ZERO Function 

Press "

ZERO

" button to enter the Zero mode, 

ZERO

 Annunciate will 

appear  and  Zero  the  display.  The  reading  is  stored  as  reference 
value for subsequent measurement.  

Press the "

 ZERO

 " button again, to exit the zero mode. 

 

5-4  BACKLIGHT Function 

When the "

" button is pressed, the backlight will be turned on.  

Press  the  button  again  to  turn  the  backlight  off.  If  the  button  is  not 
pressed  again  then  the  backlight  will  be  automatically  turned  off  in 
about 30 seconds after it is turned on. 

 

Содержание 265

Страница 1: ...Center 265 Leakage Current Clamp Meter Operating Manual ...

Страница 2: ...NTINUITY 3 3 6 RESISTANCE Ω 3 4 DESCRIPTION OF THE INSTRUMENT 4 4 1 Description of the display 4 4 2 Description of front and rear 5 5 BUTTON INSTRUCTION 6 5 1 HOLD LPF Function 6 5 2 PEAK HOLD Function 6 5 3 ZERO Function 6 5 4 BACKLIGHT Function 6 6 MEASURING INSTRUCTION 7 6 1 AC A AC mA Measurement 7 6 2 AC V Measurement 8 6 3 DC V Measurement 9 6 4 Continuity Measurement 9 6 5 Resistance Measu...

Страница 3: ...s 6 Change the battery when the symbol appears to avoid incorrect data Environmental Conditions Operation Temperature 0 C to 40 C 32 F to 104 F 80 RH Storage Temperature 10 C to 60 C 14 F to 140 F 80 RH Explanation of Symbols Attention refer to operating instructions Dangerous voltage may be present at the terminals This instrument is double insulated Approvals EN61010 300V CAT IV 2 GENERAL SPECIF...

Страница 4: ...than the maximum for the range in use the display will show Maximum Jaw Opening 23 mm Dimensions 206 x 76 x 33 5 mm Weight 262g with battery Accessories Carrying case Batteries Test Leads Instruction Manual 3 ELECTRICAL SPECIFICATION The accuracy specification is defined as percentage of reading number of counts At 23 5 C 80 RH 3 1 ACmA True RMS Range Resolution Accuracy 50 60 Hz 50 500 Hz 6 000 m...

Страница 5: ...A 2 0 5 counts 60 00 A 0 01 A 3 3 ACV True RMS Range Resolution Accuracy 50 500 Hz 60 00 V 0 01 V 1 0 3 counts 300 V 0 1 V Input impedance 1 M 3 4 DCV Range Resolution Accuracy 60 00 V 0 01 V 1 0 2 counts 300 0 V 0 1 V Input impedance 1 M 3 5 Continuity Range Buzzer Function Ohm 45 3 6 Resistance Range Resolution Accuracy 600 0 0 1 1 0 2 counts 6 000 k 0 001 k 60 00 k 0 01 k 600 0 k 0 1 k ...

Страница 6: ...n Polarity indication Low battery indication AC measurement indication DC measurement indication Current measurement indication Voltage measurement indication ZERO indication Data hold indication PEAK hold indication Low pass filter indication Continuity test indication Measurement unit Resistance measurement indication Measurement unit ...

Страница 7: ...1 Current Sensing Clamp 2 Safety protection ring 3 Backlight button 4 Clamp opening handle 5 Function select dial 6 ZERO button 7 PEAK button 8 Data hold LPF button 9 LCD display 10 COM input terminal 11 Positive input terminal 12 Battery compartment ...

Страница 8: ...old Function Selection Press PEAK button to enter the PEAK mode the meter records the peak maximum value Press the PEAK button again to exit the peak mode When you turn the rotary switch to the Ohms Continuity Test position press the PEAK button to select Resistance measurement press the PEAK button again to select continuity test with buzzer 5 3 ZERO Function Press ZERO button to enter the Zero m...

Страница 9: ...on the display Earth leakage current or any small AC current that flows through a ground wire can also be measured by this method 3 Measuring out of balance leakage current see Fig 2 Clamp onto all conductors except the ground wire The measured imbalance current value is shown on the display Before this measurement disconnect any test leads from the meter for safety In some cases wherein reading t...

Страница 10: ...o not attempt to Take any voltage measurement that may exceed this maximum to avoid Electrical shock hazard and or damage to this instrument Switch the main function selector to range Connect the red test lead to the terminal and the black lead to the COM terminal Measure the voltage by touching the test lead tips to the test circuit where the voltage measurement is required Read the result from t...

Страница 11: ... the test circuit where the value of voltage is needed Read the result from the LCD 6 4 Continuity Test With Buzzer Switch the main function to range Connect the red test lead to the terminal and the black lead to the COM terminal Connect tip of the test leads to the points between which continuity is to be measured If the resistance is under 45Ω the beeper will sound continuously Short circuit Op...

Страница 12: ...nce in a circuit make sure any power on the circuit is turned off and any capacitors are discharged 7 CHANGING THE BATTERY When the battery voltage drops below that required for proper operation of the instrument the battery symbol will appear on the LCD display and the battery should be changed Before changing the battery switch the main dial to OFF and disconnect test leads Open the back cover u...

Страница 13: ...then please remove the battery and do not store the meter in a high temperature or high humidity environment Keep the cable in the center of the clamp will return a more accurate test result when making current measurements Repairs or servicing not covered in this manual should only be performed by qualified personnel 8 2 CLEANING Periodically wipe the case with a dry cloth Do not use abrasives or...

Отзывы: