NewTom 5G
series
–
Patient Table Service Procedures
3-1
3
Patient Table with stretcher
3.1
Service Level Procedures
Special keys combination of the table console, allow to enable SERVICE level (“support modality”)
DANGER:
The usage of the SERVICE level by not authorized
personnel can result in mechanical damages of the
table and/or the scanner and injury of the user.
The SERVICE level must be used without any patient on
the table.
WARNING:
Pay attention while moving the patient's table
in order to avoid possible collision with object
and/or person.
WARNING:
Every movement can be stopped by pressing the
emergency button located under the console keypad
only if the patient table is powered through the 5G
series gantry
3.1.1
Service Level Access
From the INFO page hold the
button and press the
button. The SERVICE Level modality will be
enabled.
In order to disable the SERVICE Level modality repeat the same sequence or turn off and on the device.
INFO page
WARNING:
If the service mode is enabled, all buttons are available
in the console including longitudinal movements
(
FORWARD/BACK)
Once the SERVICE Level is enabled on the
POS
page
, press:
-
F1 to enter in the “TOOLS” page
-
F2 to enter in the “DATA DEBUG” page
-
MODE to return in the “INFO” page.
POS page