Patient Table Service Procedures
3-3
The provided actions are:
DISABLE COLLISION DETECTION
(ATTENTION: risk of collision between mechanical parts!)
Disable the anticollision security. This settings stays enabled until the device is turned off. To restore the
functionality refer to the command enable collision detection.
ENABLE COLLISION DETECTION
Enable the anticollision security (always enabled by default after device has been turned on).
COMPLETE BURNIN:
(ATTENTION: risk of collision between mechanical parts!)
Perform 40 cycles of predefined movements. To stop the process keep pressed any button for a few
seconds until the DONE.
BURNIN W/O TABLE:
(ATTENTION: risk of collision between mechanical parts!)
Perform 20 cycles of predefined longitudinal and vertical pantograph movements.
To stop the process keep pressed any button for a few seconds until the DONE.
CONSOLLE RIGHT:
Store the console position as at the patient's right.
CONSOLLE LEFT:
Store the console position as at the patient's left.
SET SERIAL NUMBER:
Set the table serial number. The serial number can be adjusted by pressing the “+” and “-” buttons.
Change digit by the arrows keys.
NEW TABLE POT. TYPE:
To setup only in case the small potentiometer cod. 97643500 is mounted (enabled by default).
Perform only for patient table with S/N:
0005 , 0012 and subsequent
OLD TABLE POT. TYPE:
To setup only in case the big potentiometer 97643400 is mounted.
Perform only for patient table with S/N:
0001 ÷ 0004 , 0006 ÷ 0011 (included)
TABLE POT. CALIB:
Perform the stretcher potentiometer calibration.
Slide all the way out the stretcher and press the MODE button. Once the procedure is succesfully carried
out the sign DONE sould appear.
XY CALIBRATION:
(ATTENTION: risk of collision between mechanical parts!)
XY axis board potentiometer calibration. To be performed only after replacement of the XY motor board
and/or potentiometer. To perform also during the table installation
X CALIBRATION:
(ATTENTION: risk of collision between mechanical parts!)
X axis potentiometer calibration. To be performed only after replacement of the X axis potentiometer.
It is part of the XY CALIBRATION process
PANT. FC BOTTOM
(ATTENTION: risk of collision between mechanical parts!)
Search for the minimum limit switch of the pantograph. Procedure performed during device
manufacturing. It is part of the PANT FC BOTH procedure.
PANT. FC TOP
(ATTENTION: risk of collision between mechanical parts!)