NewTom 5G
series
–
Patient Table Service Procedures
3-5
RESET HANNING BOARDS:
This procedure delete the firmware installed on the Hanning boards (97661177 pantograph board and
97661178 table board).
To reset the boards:
8)
Disconnect C6 connector (5G - patient table)
9)
Disconnect J8 connector located on 97660897 board.
10)
Disconnect X207 connector located on 97661178 board if you want to reset 97661177 board OR
Disconnect X207 connector located on 97661177 board if you want to reset 97661178 board.
11)
Select “RESET HANNING BOARDS” command from the service menu, insert the password and press
OK to confirm.
12)
Turn off the machine.
13)
Connect C6 connector (5G - patient table), J8 connector and X207 connector.
14)
Turn on the machine.
EXIT:
Exit from the ACTIONS page and return to the TOOLS page.
If the action task is carried out, “
DONE
” message will appear for a few seconds before returning to the
ACTIONS page.
3.1.2.2
LIMIT SWITCH page
In the
LIMIT SWITCH page
are shown the status of the hardware limit switch for the X,Y,Z,LT,UT, A (anti-
pitch safety switches and of the hand crashing security) and E (emergency button).
A white / black small circle shows the active signals. By pressing the MODE button is possible to return to
the TOOLS page.
LIMIT SWITCH page