NewTom 5G
series
–
Patient Table Service Procedures
2-4
Pressing the F1 button will confirm the request, pressing the F2 button will cancel the request and will
return to the
ACTIONS page,
pressing the MODE button will cancel the request and will return to the
HOME page
.
After confirming the selected action, in case the action required movements or a sequence that takes a
few seconds, at the end will automatically return to the HOME page, while if the action task can be
carried out instantly the “
DONE
” message will appear for a few seconds before returning to the TOOLS
page.
2.1.2.2
LIMIT SWITCH page
In the
LIMIT SWITCH page
are shown the status of the Hardware limit switch for the X,Y,Z,L,B of the anti-
pitch safety switches and of the hand crashing security and of the emergency button. An X sign shows the
active signals. In the bottom right corner the software version is shown.
By pressing the MODE button is possible to return to the TOOLS page.
LIMIT SWITCH page
X = X Axis (gantry in / out)
Y = Patient's left / right
Z = Pantograph
L = Legrest
B = Backrest
A = Anti-pitch safety switch status
E = Emergency button switch status
2.1.3
DATA DEBUG page
In the
DATA DEBUG page
appear data about the system status.
DATA DEBUG page
St
represents the CPU status. It is 0 if OK.
T
represents the elapsed seconds from the device start up.
E0
,
E1
and
E2
contains the last three error codes and
T0
,
T1
and
T2
are the time corresponding
to each error.
The D0..D3 value can assume different meaning depending on the firmware. For version equal or more
recent than 0.8 they represents:
-
D0
: pantograph potentiometer value corresponding to minimum limit switch.
-
D1
: 0 if the CAN communication is OK.
-
D2
: pantograph potentiometer value corresponding to maximum limit switch.
-
D3
: pantograph potentiometer current value
Pressing the MODE button is possible to return to the POS page.