CDN SD2204 Скачать руководство пользователя страница 4

 

Calibration

 

As the gravity value is not the same at every location on earth. To achieve best accuracy, the scale is 
highly recommended to be calibrated in compliance with the gravity at the place of use. This calibration 
process must be carried out for the first time of use, after each change of location as well as in case of 
environment temperature change. To receive accurate measuring values, it is also recommended to 
calibrate the scale periodically in weighing operation.   
 
The calibration should be made with the recommended calibration weigh

t

 (see specification).

 

Procedure for calibration: 

 

Observe stable environmental conditions. A warming up time of 30 minutes is required for stabilization.   

► Turn on the scale by pressing the

 ON/OFF

 key. 

 

 

► Press 

SET/TARE

 key and keep it pressed, until [

CAL]

 appears in the display for few seconds. 

After that the adjustment weight (see specification) appears and flashes in the display.

 

 

► Place the adjustment weight on the center of the platform. 

 

 

► Now press the 

SET/TARE

 key. Short time later [

CAL F]

 appears, then returns automatically to 

the normal weighing mode. Readings of the adjustment weight appear in the display. 

 

 

    An error during adjustment or the use of an incorrect adjustment weight will result in an error   
    message 

[

CAL E]

. Repeat adjustment. 

 

 

Keep the adjustment weight close to the scale. Daily control of the weighing exactness is recommended 
for quality-relevant applications.   
 
 

Auto-Off 

For battery operation the scale has an automatic switch-off function which can be activated or 
deactivated in the menu. Procedures as follows:   
 
Turn on scale using the 

ON/OFF

 key and wait for the zero display.

 

 

Press and hold the 

FUNCTION 

key until

 [UNIT]

 appears on the display.   

 
Press the 

UNIT/COUNT

 key or 

HOLD/PRINT

 key repeatedly until 

[ AF ]

 appears.   

 
Confirm by pressing the 

SET/TARE 

key. 

 
Use the 

UNIT/COUNT

 key or 

HOLD/PRINT

 key to choose between the following settings:

 

1.  [AF on]:

 To save battery power, the scale switches off automatically if 3 minutes without 

weighing.

 

2. 

[AF off]:

 Switch-off function deactivated.   

 
Use the 

SET/TARE

 key to confirm selected setting. The scale returns to weighing mode. 

 
 

 

 

Содержание SD2204

Страница 1: ...1 PORTION CONTROL SCALE 22 pounds 352 ounces 10 kilograms by grams ounces troy ounces pounds Hong Kong tael Taiwan tael grains pennyweights mommes tolas carats and free factor units...

Страница 2: ...amic weighing In the event that small quantities are removed or added to the material to be weighed incorrect weighing results can be displayed due to the stability compensation in the scale Example s...

Страница 3: ...rox 2 hours at room temperature after it has been disconnected from the mains Avoid static charging of the material to be weighed Major display deviations incorrect weighing results are possible if el...

Страница 4: ...enter of the platform Now press the SET TARE key Short time later CAL F appears then returns automatically to the normal weighing mode Readings of the adjustment weight appear in the display An error...

Страница 5: ...ttings in the menu ZERO Zero scale 1 Weighing Mode Turn on the scale using the ON OFF key The scale will perform self test for few seconds When the zero display appears the scale is ready for operatio...

Страница 6: ...unction can be used for unit weight control fabrication control etc Turn on scale using the ON OFF key and wait for the zero display Put the target weight on the platform and tare to zero by pressing...

Страница 7: ...RE key Weigh component Press the SET TARE key Net total sum weight of components will be displayed If necessary repeat above procedures to complete the formula up to the desired final value Back to we...

Страница 8: ...tor FFA xx xx In order to enter an own conversion factor press the UNIT COUNT key or HOLD PRINT key as explained above until FFA is displayed Press the SET TARE key to confirm the selection The last d...

Страница 9: ...ner on the scale When application involves small variations of weight it is advisable to switch off this function However the weight reading will look less stable In weighing mode press and hold the F...

Страница 10: ...RINT key several times until Pr is displayed Press SET TARE key to display current setting Press the UNIT COUNT or HOLD PRINT key to select setting based on below table Display Function Pr PC Data out...

Страница 11: ...2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 M S N1 N2 N3 N4 N5 N6 N7 N8 N9 N10 B U1 U2 U3 CR LF d Format for error condition 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 B B B B B B B B B B B E r r o...

Страница 12: ...and corrode the electronic circuits Do not subject the scale to excessive force shock dust temperature or humidity which may result in malfunction shorter electronic life span damaged battery and or d...

Страница 13: ...pon receipt of the unit prepaid at CDN PO Box 10947 Portland OR 97296 0947 This warranty does not cover damage in shipment or failure caused by failure to adhere to the accompanying instructions inade...

Отзывы: