S: Strobe Light Width Setting Command
Name
Strobe Light Width Setting command
Header
code
S
Function
It sets the pulse width for strobe light.
@ 99
S
99999 R/W FF CR
LF
Command
Header Channel
Strobe
light width
Write flag Checksum Delimiter
Response
@
99
O
FF CR LF
OK
Command Header Channel Checksum
Delimiter
@ 99
N 9
FF CR LF
NG
Command Header Channel Error No. Checksum
Delimiter
Strobe light width may be set from 20
µ
s to 33,000
µ
s.
Although it can be set increment of 1
µ
s, it will actually be rounded as shown at
“2-4 Strobe Light”. (Strobe light width can not be set with 1
µ
s increment)
F: Light Pulse Width Setting Command
Name
Light Pulse Width Setting Command
Header
code
F
Function
It sets the duty ratio for pulsed light width.
@ 99
F 999 R/W FF
CR
LF
Command
Header Channel
Light pulse
width
Write flag
Check
sum
Delimiter
Response
@ 99
O FF CR LF
OK
command
Header Channel Checksum
Delimiter
@ 99
N 9
FF CR LF
NG
command
Header Channel
Error
No.
Check
sum
Delimiter
Light intensity date 0 to 255; the pulse duty ratio varies within 0 to 100% expressed in incremental
value of100% divided by 256.
In case of Strobe Mode, the value halves to 128 increments. (0,1,3,5,7,….,255)