28 NoRAMAVAIL
The current setup requires more RAM than is currently
installed. Contact your dealer or the manufacturer.
29 PIN error
This message will appear on power-up or setup if the E
2
is
corrupted in the PIN section. Check E
2
for problems. The
access code is defaulted to the manufacturer (GSE) access
code. Also check Error 11.
Calibration Error Messages
Error Description
30 F.S.>MAX!
The entered calibration weight, together with the currently applied
signal, indicates that the full scale signal will be greater than the
allowed maximum of the indicator. Verify that correct entries have
been made for the capacity, P110, and for the calibration weight.
If all appears correct, refer to the use of the information parameter
P61100, and determine the output (in mv / volt) of the connected
load cell.
31 F.S.<.1mVv
32 ADD MORE!
The entered calibration weight, together with the currently applied
signal, indicates that the full scale signal will be less than the
allowed minimum of the indicator. Verify the proper entries for the
capacity, P110, and for the calibration weight. If all appears
correct, refer to the use of the information parameter, P61100 and
determine the output (in mv / volt) of the connected load cell. The
applied weight during calibration was less than 0.1% of capacity.
More weight than this is required. Refer to P61100 if this is
incorrect.
33 ReCALReq'd
The just completed calibration may not yield accurate results due
to either the cal weight being less than 5% of capacity or this was
the first calibration of this platform to this indicator and therefore
the coarse gain was adjusted by the indicator.
34 RES> 25K!
The current combination of capacity P110 and increment P111
result in a resolution greater than 25,000 graduations. This is
simply a warning in case this was not intended.
Error Description
35 RES>100K!
The current combination of capacity P110 and increment P111
result in a resolution greater than 100,000 graduations. This