13 OF 20
Accessing the Custom Setup List
DO NOT ATTEMPT TO ACCESS THE CUSTOM SETUP LIST DURING
CRITICAL WEIGHT PROCESSING! It is important to note that all functions of
the operating mode will be suspended immediately upon accessing the
information parameters. This includes suspension of weight conversions,
deactivation of all setpoints and cancellation of custom transmits.
The “Custom Setup” list may be accessed from the Weigh Mode. An access
code is not required to view this list.
To access the custom setup list:
1
From the Weigh Mode, key in
60205 [SELECT]
.
2
The Custom Setup list begins scrolling through each parameter to check.
If there
are no parameters to check,
Std. Setup
is displayed
3
The Custom Setup list may be repeated by pressing
[ENTER]
at P60205.
4 Press
[ZERO]
to return to the Weigh Mode.
A setup parameter that appears in the “Custom Setup” does not imply that it is
improperly configured. Consider the application and refer to the 460 Series
Technical Reference Manual for more details.
Additional Conformance Considerations
Several parameters must be considered on an individual basis as their
configuration may vary with different applications. Refer to the 460 Series
Technical Reference Manual for details.
Sealing and Audit Trails
Most legal-for-trade installations will require the indicator to be sealed. A sealed
indicator cannot be accessed for setup or calibration changes without breaking a
physical seal or incrementing an event counter, thus providing evidence that the
unit has been tampered with. Each member of the 460 Series has two types of
sealing provisions:
•
Physical seal used in conjunction with an internal program jumper
•
Three-event audit trail counter
Check with your local weights and measures authority to determine which
method(s) are required.