27
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
Note:
Outgassed byproducts may be hazardous to or noxious for operating personnel. Vacuum
pump exhaust should be vented to a location outside the workspace in a safe manner in
accordance with all applicable laws, ordinances, and regulations.
V
ENTING
P
UMP
E
XHAUST
If a pump has been installed in the cabinet,
it must be vented outside of the oven
.
Failure to do so will result in outgassed byproducts coating oven electrical system.
Additionally, this failure will expose oven operators to the outgassed byproducts.
Electronics Cabinet
The cabinet housing the vacuum pump contains exposed high-voltage electronics
and should only be accessed by a qualified electrical technician. The oven must be
disconnected from its power source when a technician is working in the cabinet.
Exhaust Lines
An exhaust line may run out either of the cabinet access ports.
Left Access Port
Right Access Port
Содержание CVAC-9.3
Страница 1: ...Installation Operation Manual CVAC 9 3 Vacuum Oven ...
Страница 5: ...5 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS ...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION ...
Страница 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Страница 34: ...34 P a g e cascadesciences com Tel 503 847 9047 CONTROL OVERVIEW ...
Страница 46: ...46 P a g e cascadesciences com Tel 503 847 9047 OPERATION ...
Страница 54: ......