OPERATION
Heating Control
The oven temperature controller stores a user-selected constant temperature set point. When
powered, the oven heats the chamber shelves to the set point. The controller board is wired to a
solid-state temperature probe located in the chamber on the rear wall. When the controller detects
that the shelf temperature has dropped below the temperature set point, it pulses power to the
heating elements.
The controller employs proportional-integral-derivative analytical feedback-loop functions when
measuring and controlling the shelving temperature. PID-controlled heating pulse intensities and
lengths are proportional to the difference between the measured shelf temperature and the current
set point. The frequency of pulses is derived from the rate of change in that difference. The integral
function slows the rate of pulses when the temperature nears the set point to avoid overshooting.
CVAC ovens rely on natural heat radiation for cooling. The oven can achieve a low-end operating
temperature of the ambient room temperature plus the oven waste heat.
The Over Temperature Limit System
When set, the mechanical OTL heating cutoff system prevents runaway heating in the oven chamber.
The OTL cutoff operates independently of the main controller and is provided with a separate,
hydrostatic temperature sensor probe located in the oven chamber. In the event the chamber
shelving temperature exceeds the current OTL setting, the OTL routes power away from the heating
elements.
The high limit is set by the end-user to a minimum of 10˚C above the highest temperature of the application
process the oven is currently being used for. Failure to set the high limit control system voids the oven
manufacturing defect warranty in the event of an overtemperature event.
After the OTL is tripped, the temperature must drop below the current OTL setting and the red Reset
button must be pushed in order to restore power to the temperature controller and heating
elements. If the temperature controller screen is dark but the vacuum display is still powered, the
OTL may have tripped.
Содержание CVAC-9.3
Страница 1: ...Installation Operation Manual CVAC 9 3 Vacuum Oven ...
Страница 5: ...5 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS ...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION ...
Страница 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Страница 34: ...34 P a g e cascadesciences com Tel 503 847 9047 CONTROL OVERVIEW ...
Страница 46: ...46 P a g e cascadesciences com Tel 503 847 9047 OPERATION ...
Страница 54: ......