OPERATION
OTL Activation During Normal Operations
Possible causes:
•
A user has set the Over Temperature Limit below the current set point for either an active
heating profile or single constant-temperature set point.
•
An external heat source or heat source inside the chamber is causing the chamber
temperature to spike.
•
The temperature controller or its sensor probe have failed and must be replaced in order to
maintain safe operation of the oven.
If the OTL activated during normal operations, adjust the control dial clockwise halfway to the next
dot to increase the setting. Then restore power to the controller by pushing the red Reset button. If
the OTL continues to interrupt heating of the oven chamber and there are no obvious external
sources of nearby heating (autoclaves, another oven), depower the oven and allow the oven
chamber to cool before opening the oven door or troubleshooting.
End of procedure
Содержание CVAC-9.3
Страница 1: ...Installation Operation Manual CVAC 9 3 Vacuum Oven ...
Страница 5: ...5 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS ...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION ...
Страница 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Страница 34: ...34 P a g e cascadesciences com Tel 503 847 9047 CONTROL OVERVIEW ...
Страница 46: ...46 P a g e cascadesciences com Tel 503 847 9047 OPERATION ...
Страница 54: ......