Setting output current
You can set output current in F61 and F62 of Set mode.
The setting range is 0.000
㎃
to 24.000
㎃
, by steps of 0.001
㎃
.
- Max. capacity : 100
㎏
, Min. capacity : 0.05
㎏
case F61 : 4000, F62 : 20000
①
case F61 : 20000, F62 : 4000
②
▣
How to use current to voltage
If F61 is set to 4mA, F62 is set to 20mA in SET mode and resistor is 250
Ω
,
the output will be 1V ~ 5V.
Note. You have to use high capacity of resistor. If you add resistor 500
Ω
,
W=I
2
R=(0.02)
2
x 500 = 0.2W
Therefore, you have to use 1/2W or more capacity and low temperature coefficient
resistor.