24
10. Calibration Mode
1) How to enter calibration mode
Remove the bolt on the rear panel and put the switch to CAL ON as figure.
①
Adjusted to CAL ON at factory.
※
Turn
②
on the power while pressing the
.
Calibrate the scale.
③
④
Put the switch to CAL OFF and perform the sealing. (See page 54)
2) Available keys
key : used for moving to the next calibration menu.
~
key : used for changing preset value.
3) Calibration menu(CAL1~CAL7)
CAL 1 : Maximum capacity set
CAL 2 : Minimum division Set
CAL 3 : Setting Weight In Span Calibration
CAL 4 : Zero calibration
CAL 5 : Span calibration
CAL 6 : Check if the calibration is done properly
CAL 7 : Input weight constant calibration after selecting national code