26
CAL 3
FUNCTION : Setting Weight In Span CALIBRATION
KEY
VF
DISPLAY
SUB
DISPLAY
DESCRIPTION
: save and go to next
menu
~
:
set value change
: input the point
L = 5000
L = 500
CAL3 SPAn
5000 kg
500 kg
Note 1. The setting weight should be within the 10 % to 100 % of maximum weight.
Note 2. If the setting weight is under the 10% of the maximum capacity, the display shows
error message, CH 12.
Note 3. If the setting weight is over the maximum capacity, the display shows error message,
CH 12.
CAL 4
FUNCTION : Zero Calibration
KEY
VF
DISPLAY
SUB
DISPLAY
DESCRIPTION
: zero calibration.
: exit
UnLOAd
- - - -
SUCCESS
CAL4 ZEro
Unload the tray and press SET
under zero calibration.
Zero calibration is completed.
The program moves into Span
calibration automatically.
Note 1. If Zero calibration is done without any error, SUCCESS message is displayed and
program moves into CAL 5 automatically.
Note 2. If the zero value is too low/high check message (“CH 14”) is displayed.
Note 3. Zero calibration can be done independently, Pressing "ZERO" key instead of “CLR
key” will perform this function.