
17
/
44
5-5. Function List detailed information.
Set-Up / Calibration Mode Selection
F00
Clear
Set-Up mode
Enter
Calibration Mode
Decimal Point Setting
F01
●
0
No Decimal point
1
1
st
place under Zero (0.0)
2
2
nd
place under Zero (0.00)
3
3
rd
place under Zero (0.000)
Back up mode selection
F02
●
0
Normal mode
1
Back up mode
Motion Band Range setting
F03
5
0
∫
9
This is set “Steady” acceptable range of weighing part.
If there is vibration on weighing part, you can set this function and
reduce the vibration effect on weighing process.
0 : Weak vibration
∫
3 : Strong Vibration
Zero Tracking Compensation Range setting
F04
5
0
∫
9
Due to external causes(Temperature, wind, and dust), there are small
weight difference, indicator will ignore the weight difference and
display Zero.
For this compensation function, indicator will estimate the weight
difference is over the set range during fixed time period.
If there is large weight difference over set range within fixed time
period, the “Zero” is breaking and will find new zero point.
Auto Zero Range setting
F05
00
00
∫
99
Within the “Auto Zero” range, weighing part is steady, indicator will
display current weight as “Zero”
If the weighing part is not “Steady”, indicator will display current
weight.
(Auto Zero Range : ± Set value + weight unit)
Digital Filter setting
F06
11
1
∫
49
Small set value for weak vibration
Large set value for strong vibration
Small
set
value
more
sensitive