44
Step 3. Max. Capacity Setting (
display)
Input Max. Capacity of Scale with No. keys.
Input Capacity and press
ENTER KEY
, and move to next step.
※
Caution
(Max. capacity value / division value)
can not be over
20,000
.(as Indicator resolution is 1/200,00).
Step 4. Zero Balance setting (
display)
Make empty the scale part, and press
ENTER KEY
.
Indicator check the current Zero balance and save the value and move next step.
Step 5. Span value calculation (
)
Input prepared Test weight value with No. keys.
And press
ENTER KEY
.
Then, display will show
and then, load prepared test weight unit on the scale.
After a few seconds(to remove the vibration effect), press
ENTER KEY
.
Then, indicator will calculate Span value and move the next step.
※
Caution
For the precise Span calibration, please prepare Test weight unit, at least 10% of Max. capacity
of Scale.
Step 6. Check Span Value and Finish the Calibration(
)
Check the Calculated Span value.
And after 3sec,
will displayed automatically and move to weighing Mode.
CAPA
XXXXXX
CAL _0_
SPAN
XXXXX
C_UP_
C_UP_
XXXXXX
C-END
C-END