9
TEST 5
FUNCTION : Key Test
KEY
LCD SCREEN
DESCRIPTION
HOLD
KEY or
kg
◀▶
lb
KEY
: execute test 5
PUSH
ZERO
Display TEST 5 state
Pressing the key to test shows
(Power, Zero, Tare, HoLd)
4. CALIBRATION MODE
How to enter calibration mode
1.
Remove the sealing bolt on the rear panel and put the switch to CAL ON.
2.
Turn on the power.
3.
You can see ‘CZ ST’ message.
4.
Put the switch to CAL OFF and perform the sealing.
Calibration menu
CZ ST : Calibration & Set mode select
UNIT
: Weight Unit Set
ZERO
: Zero Calibration
MIN
: Minimum Division Set
MAX
: Maximum Capacity Set
BAL
: Setting the Weight in Span Calibration
CAL0 ~ 3
: Span Calibration
CZ ST
Function : Calibration mode & Set mode select
KEY
LCD SCREEN
DESCRIPTION
ZERO :
Operate the Calibration
mode
TARE :
Operate the Set mode
CZ ST
Содержание NC-1
Страница 1: ...2 ...
Страница 19: ...20 9 Click Reset button Then you can see message like this dialog box 10 Click Download button ...
Страница 20: ...21 11 Select hex file 12 Download is done 13 If download is done you must turn off SW5 FIG 2 ...
Страница 24: ...4 Cal schematic SCHEMATIC DIAGRAM 25 ...
Страница 25: ...26 8 PCB BOARD DIAGRAM 1 Main PCB board LED ...
Страница 26: ...27 2 Display PCB board ...
Страница 27: ...28 3 Charger PCB board ...
Страница 28: ...29 9 DEVICE OF SPECIFICATION 1 MPU ADUC845 ...
Страница 29: ...30 ...
Страница 30: ...31 ...
Страница 31: ...32 ...
Страница 32: ...33 2 OPAMP OP2177 OP2277 ...
Страница 33: ...34 ...
Страница 34: ...35 ...
Страница 35: ...36 3 REGULATOR LM2575 ...
Страница 36: ...37 ...
Страница 37: ...38 ...
Страница 38: ...39 ...
Страница 39: ...40 ...
Страница 40: ...41 ...
Страница 41: ...42 ...
Страница 42: ...43 4 RESET TC1274 ...
Страница 43: ...44 ...
Страница 44: ...45 ...
Страница 45: ...46 ...
Страница 46: ...47 4 RESET TC1274 ...
Страница 47: ...48 ...
Страница 48: ...49 ...
Страница 82: ...83 12 EXPLODE VIEW ...
Страница 83: ...84 ...