51
Err 11 (at Span calibration)
■
Cause
The weight for span calibration is greater than 100% of the maximum capacity.
■
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
Err 12 (at Span calibration)
■
Cause
The weight for span calibration is lower than 10% of the maximum capacity.
■
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
Err 13 (Zero Range when the Power is ON)
■
Cause
The initial zero range exceeds above 10% of maximum capacity
■
Management
Unload the scale
Check the A/D value (at ‘TEST 3’).
If the load cell is damaged, replace the load cell
Err 14 (at Span calibration)
■
Cause
The weight for span calibration is 0kg.
■
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
BAT or LobAt
Содержание NC-1
Страница 1: ...2 ...
Страница 19: ...20 9 Click Reset button Then you can see message like this dialog box 10 Click Download button ...
Страница 20: ...21 11 Select hex file 12 Download is done 13 If download is done you must turn off SW5 FIG 2 ...
Страница 24: ...4 Cal schematic SCHEMATIC DIAGRAM 25 ...
Страница 25: ...26 8 PCB BOARD DIAGRAM 1 Main PCB board LED ...
Страница 26: ...27 2 Display PCB board ...
Страница 27: ...28 3 Charger PCB board ...
Страница 28: ...29 9 DEVICE OF SPECIFICATION 1 MPU ADUC845 ...
Страница 29: ...30 ...
Страница 30: ...31 ...
Страница 31: ...32 ...
Страница 32: ...33 2 OPAMP OP2177 OP2277 ...
Страница 33: ...34 ...
Страница 34: ...35 ...
Страница 35: ...36 3 REGULATOR LM2575 ...
Страница 36: ...37 ...
Страница 37: ...38 ...
Страница 38: ...39 ...
Страница 39: ...40 ...
Страница 40: ...41 ...
Страница 41: ...42 ...
Страница 42: ...43 4 RESET TC1274 ...
Страница 43: ...44 ...
Страница 44: ...45 ...
Страница 45: ...46 ...
Страница 46: ...47 4 RESET TC1274 ...
Страница 47: ...48 ...
Страница 48: ...49 ...
Страница 82: ...83 12 EXPLODE VIEW ...
Страница 83: ...84 ...