Menu-F3.06: Maximum weight of A-out
Function
Maximum Output Weight Value upon Using Analog Out option
Display Part
Meaning
Set Range
(0 ~ Max weight)
1000
2000
Initial value : Max weight
Maximum output at 1000 kg
Maximum output at 2000 kg
Menu-F3.07: Adjust zero of A-out
Function
Adjust the Zero Output upon Using Analog Out option
Display Part
Meaning
0000
0.000mA, 0V output
4000
4.000mA, 2V output
Set Range
(0 ~ 24000)
4015
4.015mA, 2.007V output
Reference 1. The example is based on settings of 0~10V, 0~24mA.
Reference 2. The Micro-adjustment function does not work during simultaneous output of V-out and I-out.
Menu-F3.08: Adjust Span of A-out
Function
Adjust the Maximum Output upon Using Analog Out option
Display Part
Meaning
10000
10.000 mA, 4.16 V output
20000
20.000 mA, 8.33 V output
Set Range
(0 ~ 24000)
24000
24.000 mA, 10.0 V output
Reference 1. The example is based on settings of 0~10V, 0~24mA.
Reference 2. The Micro-adjustment function does not work during simultaneous output of V-out and I-out.
Содержание EXI-200AD
Страница 1: ... 2 이미지 추가 ...
Страница 20: ...21 4 External Explanation 4 1 External Specifications ...
Страница 28: ...29 2 DCType ...
Страница 29: ... 30 5 1 2 Load Cell Connector ...
Страница 31: ... 32 5 1 3 RS 232 RS 485 Connector RS 485 RS 232 COM2 RS 232 COM1 ...
Страница 86: ...87 Weight Date 8 byte Byte No Example weight 1 2 3 4 5 6 7 8 13 5kg 1 3 5 135kg 1 3 5 135kg 1 3 5 ...
Страница 98: ...99 MEMO ...
Страница 99: ... 100 MEMO ...
Страница 100: ...101 ...