6.3.5 CAL 5 (Direct Calibration))
CAL 5-1
Function: Direct input about the zero value of loadcell
Range of set value: 1 ~ 99,999
Used key
Display
Descriptions
ZE-CAL
Direct zero input mode
0.0000
Loadcell zero = 0.0000 mV/V
Using numeric keys
~
Enter the output value.
=Set,
=Cancel
0.1000
Loadcell zero = 0.1000 mV/V
CAL 5-2
Function: Direct input about the span value of loadcell
Range of set value: 1 ~ 99,999
Used key
Display
Descriptions
SP-CAL
Direct span input mode
1.2000
Loadcell span = 1.2000 mV/V
Using numeric keys
~
Enter the output value.
=Set,
=Cancel
2.0000
Loadcell span = 2.0000 mV/V