9.4 Analogue Output Setting
Menu-F3.01: A-out activation range setting
Function
Set A-Out range
Display Part
Meaning
3-01. 0
Unipolar(operating in +weight)
Set Range
(0, 1)
3-01. 1
Bipolar(operating in
±
weight)
Menu-F3.02: V-out range setting
Function
V-out range
Display Part
Meaning
3-02. 0
V-out is closed
3-02. 1
0V ~ 5 V
3-02. 2
0V ~ 10 V
3-02. 3
±5 V
Set Range
(0~4)
3-02. 4
±10 V
Reference 1. Simultaneous output is not used. Therefore the V-out output range settings are applied with
priority during setting.
Reference 2. When only using I-out, the V-out output range must be set to 0.
Menu-F3.03: I-out range setting
Function
I-out range
Display Part
Meaning
3-03. 0
I-out is closed
3-03. 1
4 mA ~ 20 mA
3-03. 2
0 mA ~ 20 mA
Set Range
(0 ~ 3)
3-03. 3
0 mA ~ 24 mA
Menu-F3.04: Dual output(V-out & I-out)
Function
Dual output(V-out & I-out)
Display Part
Meaning
3-04. 0
Dual output disabled
Set Range
(0, 1)
3-04. 1
Dual output enabled
Menu-F3.05: Minimum weight of A-out
Function
Minimum Output Weight Value upon Using Analog Out option
Display Part
Meaning
Set Range
(0 ~ Max weight)
1000
2000
Initial value : 0
Maximum output at 1000 kg
Maximum output at 2000 kg
Содержание EXI-200AD
Страница 1: ... 2 이미지 추가 ...
Страница 20: ...21 4 External Explanation 4 1 External Specifications ...
Страница 28: ...29 2 DCType ...
Страница 29: ... 30 5 1 2 Load Cell Connector ...
Страница 31: ... 32 5 1 3 RS 232 RS 485 Connector RS 485 RS 232 COM2 RS 232 COM1 ...
Страница 86: ...87 Weight Date 8 byte Byte No Example weight 1 2 3 4 5 6 7 8 13 5kg 1 3 5 135kg 1 3 5 135kg 1 3 5 ...
Страница 98: ...99 MEMO ...
Страница 99: ... 100 MEMO ...
Страница 100: ...101 ...